12. Troubleshooting
52
SOLUTION:
•
The thermopile in your chamber, which measures
sample temperature, may have become contami-
nated.
Refer to Chapter 10 for directions on cleaning.
•
If you weren’t using Decagon’s KCl verification
standards,
high readings may indicate that the salt
solution you are using is not in equilibrium.
4. PROBLEM:
Message on screen displays the following (example):
SOLUTION:
•
The sample is too dry for the instrument to read
accurately.
If your sample has a water potential that is
above the detection limits of the instrument, this mes-
sage will come up. Essentially, it means that there is
not enough sample moisture to condense on the mir-
ror and provide a reading.
•
The mirror may be dirty.
Try cleaning the mirror
and chamber and measuring the sample again.
< -820 MPa