WP
4C
PotentiaMeter
12. Troubleshooting
54
SOLUTION:
•
The block is not plugged in to the motherboard. Open
the case and check to make sure that the small ribbon
cable that connects the block to the motherboard is
snapped and locked in place.
•
One or more components has failed on the block’s cir-
cuit board. If the block is properly plugged in to the
motherboard and this message appears, it is likely that
one or more of the components have failed on the
block’s circuit board.
Sensor Performance Screen
If, after cleaning your instrument and reading the other
troubleshooting hints, you have reason to believe that one
of the components of your WP4C may be causing mea-
surement error, you can access a screen that will display
values for component performance. This is done by hold-
ing down the lower right button while turning on the
instrument. After it initializes, it will beep and come to the
following screen:
This screen gives you four values. The top left value is the
value the thermocouple is reading. It is basically the differ-
-Exit-
sensors
3.36 0.030
23.5 0.665