- 11 -

DIGITAL WEIGHING INDICATOR

- 10 -

P10

Digital filtering intensity

X=1:

high

X=2:

medium

X=3:

low

P11

Stable time

X=1:

high

X=2:

medium

X=3:

low

P12

Stable scope

X=1:

low

X=2:

medium

X=3:

high

Connect Scoreboard to Indicator (Optional)

Serial Communication and Indicator Connection

1. 

Current loop or RS232 signal is used for scoreboard connection, which is transmitted in serial 

 

binary code style. The baud rate is adjustable.  

 

 

 

• Make sure that scoreboard and signal output is connected correctly. If there is something wrong 

 

with the connection, damage will happen to the output port of the instrument and input port of 

 

the scoreboard.  

 
 

• Only use the provided connecting cable

1. 

Make sure that communication interface output and computer are connected correctly. Iif there 

 

is something wrong with the connection, damage will happen to the output port of the instrument 

 

and the input port of computer. Data format for the RS232 output is the following: the 1st is 

 

starting the bit, the 10th is the stop bit, and the middle are 8 data bits.

Communication mode as follows:

1. 

In continuous mode: 

 

The data transmitted is weight (Gross weight or net weight) 

 

The format of G.W.: ww000.000kg or ww000.000lb 

 

The format of N.W: wn000.000kg or wn000.000lb 

 

Note:The position of above decimal is decided by the decimal set on the indicator. 

2. 

In command mode: 

 

The indicator performs the corresponding operation according to the command transmitted 

 

from the pc. 

 

Command R The indicator sends weight data once in receipt of command R from pc 

 

(the format is the same as the continuous mode) 

 
 

Command T The indicator tares in receipt of command T from pc (the same as tare key). 

 

If no command is received the indicator displays CR LF  

 
 

Command Z The indicator zero in receipt of command T from Z (the same as zero key If no 

 

command is received the indicator displays CR LF 

1

2

3

4

5

B port

RS 232 C 5 pin (Male)

A port

CH2.54-8P

Orange 1

Red 2

Black 3

4

5

6

7

8

Orange TXD

Black GND

3 Transmit Data
2 Receive Data
5 Chassis Ground
1 Carrier Detect
4 Data Terminal Ready
6 Data Set Ready
7 Request To Send
8 Clear To Send

9 Pin Port (Female)
Computer

5 Pin Port (Male)

RXD 2
TXD 3
GND 5

Red RXD

Summary of Contents for D1047067

Page 1: ...Digital Weighing Indicator User Manual WWW DEELAT COM ...

Page 2: ...Indication 13 Chargeable Battery 13 Maintenance 14 Notes 15 Model Deelat Floor Scale weighing indicator Accuracy GradeIII n 3000 Sample Rate 10 times second A D conversion Δ Load cell sensitivity 1 5 3mV V Scale interval 1 2 5 10 20 50 for selection Display 6 bits LCD 6 state indicating signals Scoreboard interface optional In serial output mode current loop signal transmission distance 50m Commun...

Page 3: ... come into counting mode Press this button to accumulating the weight in weighing mode Press this button for sample taking in counting mode Press this button to tare in weighing mode Press this button to zero in weighing mode Press this button to turn on off the scale Press this button to stay in weighing mode or Press this button to save move in set mode FUNC TARE ZERO Unit Kg Charge A12 COUNT FU...

Page 4: ... zero 2 If the displayed value deviates from the zero point but it is still within the manual zero range the ZERO key will return the indictor to zero If not within the manual zero range then the ZERO key will not function If this is the case please recalibrate or reset zero parameters 3 Only when a stable indication is displayed can the zero operation function 1 When the indicator is in weighing ...

Page 5: ... keep the result from the last sampling 1 In weighing mode hold and press Func for 5 seconds and it will enter user function setting mode mode P There are 12 modes from P1 to P12 Press to choose the mode and press Tare to choose the parameter The description of parameters is as follows P1 kg Lb change X 1 kg display X 2 Lb display P2 automatically power off X 1 N A X 2 Power off in 10 minutes X 3 ...

Page 6: ...e following the 1st is starting the bit the 10th is the stop bit and the middle are 8 data bits Communication mode as follows 1 In continuous mode The data transmitted is weight Gross weight or net weight The format of G W ww000 000kg or ww000 000lb The format of N W wn000 000kg or wn000 000lb Note The position of above decimal is decided by the decimal set on the indicator 2 In command mode The i...

Page 7: ...point full range and enter into the zero point calibration mode Press ZERO when the stable signal is on and it will display End and keep the original full range point calibration Press the calibration switch under the lead sealing board at the back of the indicator and it will save the setting and return to the weighing status 4 Fast calibration for full range point Fast calibration for full range...

Page 8: ...pt anti static measures 4 Do not clean the case of indicator with intensive solvents for example benzene and nitro oils 5 Liquid and conductive substances should not come into contact with the indicator otherwise the electronic components can be damaged and the user could get shocked 6 Make sure to cut off the power supply to the indicator and relevant devices before connecting and disconnecting a...