Calibration

Error Indication

Chargeable Battery

1. 

Connect the load cell properly and then turn on the indicator and press the [#] key during initial-

ization. It will enter into the calibration mode and calibrate as following:  

STEP

OPERATION

DISPLAY

NOTES

1

Press [TARE] for selection of 
division

[d X]

Select optional division (1/2/5/10/20/50), 
press [#] to confirm 
Example: 20 

2

Press [TARE] for selection of 
DECIMAL POINT selection

[P X ]

Select optional decimal point: 0~3, press [#] 
to confirm 
Example: 3 

3

Set the full range

[FULL]

Press [TARE] for selection of the digit bit; Press 
[ZERO] for selection of the digit; Press [#] to 
confirm the input of the full range 

4

Zero point calibration: Press [#] 
when the stable signal is on

[nOLOAD]

Confirm there is no load 

5

Full range point calibration: 
Press [#] when the value input 
is the same as the loaded weight 
and the stable signal is on

[AdLOAD]

While inputting the loaded weight, 
Press [TARE] for selection of the digit bit; Press 
[ZERO] for selection of the digit; when the 
input value is the same as the loaded weight 
and the digit bit is at the highest bit, press [#] 
when the stable signal is on

6

[End]

7

Press the calibration span under 
the lead sealing board at the 
back of the indicator

It saves the calibration parameter and returns 
to the weighing status. 
Attention: If you do not press the calibration 
switch the parameters will not be saved 

2. 

FAST CALIBRATION FOR ZERO POINT AND FULL RANGE POINT 

 

Press [#] while it is initializing and it will enter into the calibration mode. 

3. 

Fast calibration for zero point: 

 

At any point before it displays [nOLOAD], press [FUNC] and it will keep the original division, 

 

decimal point, full range and enter into the zero point calibration mode. Press [ZERO] when the 

 

stable signal is on and it will display [End] and keep the original full range point calibration. Press 

 

the calibration switch under the lead sealing board at the back of the indicator and it will save the 

 

setting and return to the weighing status. 

4. 

Fast calibration for full range point: 

Fast calibration for full range point: At any point before it shows [AdLOAD], press [ACCU] and it will 
keep the original division, decimal point, full range, zero point calibration and enter into the full range 
point calibration mode. When it is finished, press the calibration switch under the lead sealing board 
at the back of the indicator and it will save the setting and return to the weighing status. 

EER 1 

The AD value is too small when calibrated at full range point.

EER 2 

The zero point is out of range when calibrated at zero point.

EER 3 

The zero point is out or auto zero range upon starting

EER 4 

The input sample quantity is zero when sampling in counting mode.

EER 5 

The input weight is zero when full scale calibrated in calibrating mode.

EER 6 

The unit weight is less than 0.25e when sampling in counting mode

bAt-lo 

Low power

1. 

Turn on the AC power and the indicator will charge the battery automatically. If you don’t use the 

 

battery frequently, you should take the battery out.  

 

• Note: red end is +, and the black end is -. Incorrect connection will ruin the indicator.  

 

• Note: The built-in battery should be fully charged before it is used for the first time.  

2. 

When you turn off the AC power, and push start key the battery will work. When [LouoL] is 

 

displayed it means there is insufficient voltage and it needs to charge.  

3. 

When you use the battery for the first time, you should charge the battery for 20 hours in order 

 

to ensure maximum battery life.  

4. 

If you don’t use battery for a long time, you should charge the battery for 10-12 hours every 

 

2 month period to prolong battery life.

Summary of Contents for D1047067

Page 1: ...Digital Weighing Indicator User Manual WWW DEELAT COM ...

Page 2: ...Indication 13 Chargeable Battery 13 Maintenance 14 Notes 15 Model Deelat Floor Scale weighing indicator Accuracy GradeIII n 3000 Sample Rate 10 times second A D conversion Δ Load cell sensitivity 1 5 3mV V Scale interval 1 2 5 10 20 50 for selection Display 6 bits LCD 6 state indicating signals Scoreboard interface optional In serial output mode current loop signal transmission distance 50m Commun...

Page 3: ... come into counting mode Press this button to accumulating the weight in weighing mode Press this button for sample taking in counting mode Press this button to tare in weighing mode Press this button to zero in weighing mode Press this button to turn on off the scale Press this button to stay in weighing mode or Press this button to save move in set mode FUNC TARE ZERO Unit Kg Charge A12 COUNT FU...

Page 4: ... zero 2 If the displayed value deviates from the zero point but it is still within the manual zero range the ZERO key will return the indictor to zero If not within the manual zero range then the ZERO key will not function If this is the case please recalibrate or reset zero parameters 3 Only when a stable indication is displayed can the zero operation function 1 When the indicator is in weighing ...

Page 5: ... keep the result from the last sampling 1 In weighing mode hold and press Func for 5 seconds and it will enter user function setting mode mode P There are 12 modes from P1 to P12 Press to choose the mode and press Tare to choose the parameter The description of parameters is as follows P1 kg Lb change X 1 kg display X 2 Lb display P2 automatically power off X 1 N A X 2 Power off in 10 minutes X 3 ...

Page 6: ...e following the 1st is starting the bit the 10th is the stop bit and the middle are 8 data bits Communication mode as follows 1 In continuous mode The data transmitted is weight Gross weight or net weight The format of G W ww000 000kg or ww000 000lb The format of N W wn000 000kg or wn000 000lb Note The position of above decimal is decided by the decimal set on the indicator 2 In command mode The i...

Page 7: ...point full range and enter into the zero point calibration mode Press ZERO when the stable signal is on and it will display End and keep the original full range point calibration Press the calibration switch under the lead sealing board at the back of the indicator and it will save the setting and return to the weighing status 4 Fast calibration for full range point Fast calibration for full range...

Page 8: ...pt anti static measures 4 Do not clean the case of indicator with intensive solvents for example benzene and nitro oils 5 Liquid and conductive substances should not come into contact with the indicator otherwise the electronic components can be damaged and the user could get shocked 6 Make sure to cut off the power supply to the indicator and relevant devices before connecting and disconnecting a...

Page 9: ...WWW DEELAT COM ...

Reviews: