In Zero Span, the start frequency, stop frequency and center
frequency are always the same value. For more test examples, please
visit www.deepelec.com .
Calibration
Calibration and Normalization
The VNA Master is a field portable unit operating in the rigors of the
test environment. In order to ensure measurement accuracy, RF
calibration (SOLT) must be performed prior to making a measurement
in the field. Calibrating with the specified mechanical calibration
requires three calibration parts: open, short, and match(load) and one
RG316 RF cable. Calibration data is saved as user calibration data. It
can be saved to status 0/1/2/3/4, which will be automatically loaded at
the next boot and can be also be loaded via the RECALL menu.
The figure below shows the OPEN, SHORT, and LOAD calibration
components that are standard with the DeepVNA 101.
Brief summary of DeepVNA 101 calibration steps:
1) Open
CAL
menu and press
RESET
2) Open
CALIBRATE
menu
3) Connect SMA Open to PORT1 wait more than 3 seconds (wait for the
screen to stabilize) and press
OPEN
4) Connect SMA Short to PORT1 wait more than 3 seconds (wait for the
screen to stabilize) and press
SHORT
5) Connect SMA Load to PORT1 wait more than 3 seconds (wait for the
screen to stabilize) and press
LOAD
6) Connect SMA Load to PORT1, connect second Load to PORT2, wait
more than 3 seconds (wait for the screen to stabilize) and press
ISOLN
Note:
if you don't have second Load, just leave PORT2 open.
7) Connect RF cable from PORT1 to PORT2 wait more than 3 seconds
(wait for the screen to stabilize) and press
THRU
8) Press
DONE
, save to 0/1/2/3/4 one of them.