Steps to Deploy the Lifeline ARM:
The following instructions are for a two-rescuer scenario that begins after it has been confirmed
that the patient is unresponsive and not breathing, clothing has been removed from the patients
chest and manual CPR has begun.
1. Open the Carrying Case and remove the Backboard.
Place the Backboard next to the patient.
2. Place the Backboard under the patient.
Lift and roll the patient as needed to slide the Backboard underneath the patient’s armpits
and in line with the nipple line according to the Patient Diagram. Accurately placing the
Backboard now will make it easier to correctly align the Compression Module later on.
Resume manual CPR.