20
enter measurement mode
10
Reserved
11
reserved
12
Error reading the internal
FLASH
Restart the detector. If the
error persists, contact the
manufacturer
13
Error reading the internal
FLASH
Restart the detector. If the
error persists, contact the
manufacturer
14
Error reading the internal
FLASH
Restart the detector. If the
error persists, contact the
manufacturer
15
Sensor error
Replace the sensor
16
The device exceeded the
maximum calibration
interval by 50 %
Calibration is necceserry
129
The measured value
exceeds the range of -10%
If the error persists,
contact the manufacturer
130
The result of the AD
covertion is out of range
Contact the manufacturer
4. Signalization transmitted by the current loop 4-20 mA
Measurement:
The measure concentration is directly proportional to 4-20 mA current output
Exceeding the range of measured concentrations:
Current output ranges from 20-22 mA
End of valid calibration:
Current output transmits the actual measured concentration for 10 s in a 4-20 mA range and
for 1 s 2 mA current
Malfunction:
Current output will be set to 0,5 mA
Mainternace: