6E-2. Monitoring Equipment
OZONE IN AIR
Electronic Ambient Ozone Monitor
a. Shall measure the level of ozone present
in the room where ozone equipment is
located.
b. The monitor shall include both visible and
audible alarms. The EcoZone™ Model
EZ-1 portable ambient ozone analyzer by
EcoSensors, Inc. is recommended for use
with the MPI-300.
b-1. EcoZone™ Model EZ-1 features:
• Constantly monitors work environment
• O
3
concentration shown by a multi-color
display from green (safe) to yellow (cau-
tion) to red (danger)
• No installation required
• Easily understood by non-technical per-
sonne
• Field work and “walk around” analyses
possible via power from internal recharge-
able batteries
• Compatible with EcoSensor VOC and O
3
sensing, monitoring, and control products
b-2. EcoZone™ Model EZ-1 specifications:
• Range: 0.02-0.14 PPM O
3
• Bar graph display: normally green, yellow
at 0.05 PPM (caution), red at 0.1 PPM
(danger)
• Response time: with a few seconds of O
3
reaching the sensor
• Accuracy: within 20% in the 0.05-0.1
PPM range
• Recommended calibration check annually
• Temperature Range: 18-37°C (65-100°F)
• Measurement principle: HMOS (Heated
Metal Oxide Semiconductor) sensor
• Size: 85 x 35 x 60 mm (3 1/4” x 1 3/8” x
2 3/8”)
• Weight: 154 grams (6 oz.)
• Power Requirements: 12-24 volts DC
at 300-500 mA. AC adapters available
worldwide.
• Batteries: self-contained NiMH. Approxi-
mately 2 hour capacity. Recharged by AC
adapter overnight.
Badge-Type Ambient Ozone Monitor
a. Shall measure the level of ozone present
in the room where ozone equipment is
located.
b. The ChromAir
®
System ozone badge (K&M
Environmental P/N 380010-10; www.
kandmenvironmental.com; 1-800-808-
2234) may be used for personnel or area
monitoring for exposure times ranging from
5 minutes to 10 hours. For higher resolution,
the ChromAir ozone badge may be used in
conjunction with the ChromAir ozone color
comparator.
b-1. ChromAir System features:
• Constantly monitors work environment
• Accurate measurements
• Ease of use - requires minimal training
• Cost effective - offers the most inexpen-
sive air sampling solution available
b-2. ChromAir System specifications:
• Color change: blue to white
• Exposure range: 0.08–1.6 PPM/hr
• Max. sampling time: 10 hours
• Min. sampling time: 5 minutes
• Relative humidity range: 20% - 90%
• Face velocity range: 10-165 cm/sec
• Temp. range: 16°C - 30°C (61°F - 86°F)
•
Light effect - sunlight: not recommended
• Light effect - visible: no effect
• Dimensions: 10.5 cm x 5.5 cm x 0.25 cm
• Weight: 11 g
• Refrigerated shelf life: 1 year
Other ambient ozone monitors are also avail-
able. Contact DEL for further information.
OZONE IN WATER
The ozone concentration in water can be measured
in different ways.
1. Oxidation Reduction Potential (ORP) monitor
measures the effective biocidal activity of
dissolved ozone in water in millivolts and can
be converted to mg/l (PPM). NOTE: ORP
is an
ineffective measurement of ozone
concentrations greater than about 0.5 PPM.
2. Dissolved Ozone monitors measure levels
of dissolved ozone in water (mg/l or PPM).
The HACH Indigo Blue AccuVac high range
dissolved ozone test kit (DEL P/N 6-0116) is an
inexpensive colorimetric test. Electronic, in-line
dissolved ozone monitors are also available.
Contact DEL for further information.
6E-3. Ozone Detection Data Log
In addition to automatic chart recorders, operators
should log ozone monitor output. Record ozone
output data twice per shift (maximum 12 hour shift).
If system does not have an automatic chart recorder,
record data every 4 hours, regardless of shift
changes.
(See Daily Data Log - page 10).
Summary of Contents for MPI-300
Page 13: ...10 MPI 300 Installation Operations Manual APPENDIX A DAILY DATA LOG...
Page 14: ...11 MPI 300 Installation Operations Manual APPENDIX A...
Page 15: ...12 MPI 300 Installation Operations Manual APPENDIX B EMERGENCY INFORMATION...
Page 16: ...13 MPI 300 Installation Operations Manual...
Page 17: ...14 MPI 300 Installation Operations Manual APPENDIX C SAFETY...