- 35 -
(equal to the number of interposed repeaters plus 1). Read-only parameter if the
instrument is connected to a base unit.
5)
RF_OFF_LINE
: enables or disables the RF stage of the data logger. Select
NO
to
activate the RF stage. Activation or deactivation of the RF stage can be done also
through the connection button.
6)
EXIT
: returns to the main menu.
4) THLD_MENU
(alarm thresholds)
1)
Quantity 1_DOWN_THLD
: lower alarm threshold of quantity 1. The type of quan-
tity depends on the data logger model.
2)
Quantity 1_UP_THLD
: higher alarm threshold of quantity 1. The type of quantity
depends on the data logger model.
3)
…
4)
Quantity n_DOWN_THLD
: lower alarm threshold of quantity n. The type of quan-
tity depends on the data logger model.
5)
Quantity n_UP_THLD
: higher alarm threshold of quantity n. The type of quantity
depends on the data logger model.
6)
THLD_ALRM
: enables or disables the buzzer when measurement thresholds are
exceeded.
7)
EXIT
: returns to the main menu.
5) MEAS_UNIT_MENU
(measurement unit)
1)
Quantity 1_UNIT_MEAS
: measurement unit of quantity 1. The type of quantity
depends on the data logger model.
2)
…
3)
Quantity n_UNIT_MEAS
: measurement unit of quantity n. The type of quantity
depends on the data logger model.
4)
EXIT
: returns to the main menu.
Note
: the unit of measurement is changed only on the LCD; the data are always sent in
the unit of measurement set in the base unit.
6) LOG_MENU
(logging)
1)
LOG_STAT
: enables or disables the logging.
2)
LOG_CYCL
: choice between cyclical management (the new data overwrite the old
ones when the memory is full) or non-cyclical management (logging stops when the
memory is full) of the data logger memory. Select
YES
for the cyclical management.
3)
LOG/RF_TIME
: choice of logging and RF transmission interval (the two intervals
coincide). If it is higher than the measuring interval, the average of the measure-
ments acquired during the interval will be stored.
4)
MEAS_TIME
: choice of the measurements acquisition interval. It is forced to the
value
Log/RF_TIME
if a higher value is set
.
5)
LOG_DEL
: deletes all stored measurements from the data logger memory. Select
YES
to delete the memory.
6)
EXIT
: returns to the main menu.
7) CLK_MENU
(clock)
1)
YEAR
: year. Read-only parameter if the instrument is connected to a base unit.
2)
MON
: month. Read-only parameter if the instrument is connected to a base unit.
3)
DAY
: day. Read-only parameter if the instrument is connected to a base unit.
4)
HOUR
: hour. Read-only parameter if the instrument is connected to a base unit.
5)
MIN
: minutes. Read-only parameter if the instrument is connected to a base unit.
6)
EXIT
: returns to the main menu.