ENGLISH
98
KEY
SIMBOLS LIT BESIDES
THE NUMBERS
FUNCTION
DESCRIPTION
With a data transmission speed of
300 baud (the character being
composed of 10 bit: 1 start bit + 8
character bit + 1 stop bit), the time
taken to send 80 characters on
the serial line is: 80 / (300/10) =
2.7 sec.
There is therefore a limit on the
minimum time that may be set as
a function of the baud rate cho-
sen:
If a time of less than 4 sec. with a
speed of 300 baud is accidentally
set during programming, the time
is automatically corrected to
4 sec.
The Serial Out/Memory symbol
flashes during the immediate
unloading of data onto the serial
line. The instrument does not
switch off automatically. The func-
tions of all the keys are disabled,
preventing the performance of all
their functions except the Serial
Out key and the ON/OFF key.
When pressed again, the Serial
Output key ends the unloading of
data in progress. A report is
issued stating the maximum, mini-
mum and mean values.
BAUD
RATE
300 baud
600 baud
> 600 baud
MINIMUM
SETTABLE
TIME
4 sec.
2 sec.
1 sec.