background image

Battery replacement

HD2013-DB   uses   an   internal  3.6 V   lithium-thionyl   chloride  (Li-SOCl

2

not   rechargeable 

battery,  size   C,   with  Molex   5264  2-pole   connector.   The   charge   status   of   the   battery   is 
constantly monitored and displayed.
When the battery voltage drops below a certain value, the instrument automatically turns off 
the backlit of the display, so as to minimize the battery consumption and allow the instrument 

to   continue   to   perform   the   measurement   functions.   When   the   backlit   is  turned   off,   it   is 
recommended to replace the battery soon.
If the battery voltage drops further to a level that does not allow the correct operation of the 
instrument, the message "CHANGE BATTERY NOW!" is displayed; in such a case, the battery 

replacement is necessary.

The instrument is  equipped  with non-volatile memory

: the data remain stored even if 

the battery is discharged and disconnected.
To replace the battery, proceed as follows:

1.

disconnect the instrument from the PC, if connected;

2.

unscrew the four screws fixing the front cover;

3.

remove the battery connector and take the battery out of its holder;

4.

insert the new battery paying attention to the correct polarity (the connector is provided 
with a protection device that prevents the reversal of polarity);
as an alternative to the lithium battery, if not available, three 1.5 V alkaline batteries can 

be  used, to  be  inserted  in the  batteries holder  shown in  the  figure. In such a case, 
connect  the  batteries  holder   connector  to   the   battery  connector   on the  board  of the 

instrument;

5.

close the cover with the fixing screws; when closing, verify that the sealing gasket is 
properly positioned; the screws must be securely fastened.

Fig. 4: Inside view of the data logger

5

Battery

connector

Reset

push-button

Lithium

battery

Batteries holder
for alkaline

batteries

Reviews: