HD32.3
- 17 -
V1.4
3.
USER
INTERFACE
The user interface is composed of an
LCD
graphic back - enlighten display
and start and set keys
of the instrument. With a battery power supplying without pressing any key, the back- lighting
switches off after 1 minute. To activate it again, press one key. With external power supply, the
back-lighting is always activated.
To switch the instrument on and off, press
ON/OFF
key. When you switch it on, the logo and the
model of the instrument will be visualized, for some seconds, then you can go to the main visualiza-
tion.
3.1
THE DISPLAY
HD32.3
visualizes the parameters for the calculation of
WBGT
index and for the calculation of
PMV-PPD
indexes. During the normal working, pressing the key
▼
you can go from the visualiza-
tion of
WBGT
index parameters to the
PMV – PPD
index parameters visualization.
Visualization of
WBGT Index:
WBGT Index
2008/11/28 08:00:00
Log 00 00:00:00
Tn 15.6 °C
Tg 20.2 °C
T 20.2 °C
WBGT(in) 17.0 °C
WBGT(out) 17.0 °C
The first line visualizes the
charge condition of the battery
, the
second line indicates the current
date and the current hour
. If logging function is activated, the third line indicates the current log-
ging number and the spent time from the logging start.
The
detected sizes
are:
Tn:
wet bulb temperature with natural ventilation
Tg:
globe thermometer temperature
T:
ambient temperature
WBGT (in):
WBGT index in absence of direct solar irradiation
WBGT (out):
WBGT index in presence of direct solar irradiation
Pressing the key
▼
you can visualize the
PMV-PPD index
:
PMV Index
2008/11/28 08:00:00
Log 00 00:00:00
Va 0.00 m/s
Tg 22.0 °C
T 22.0 °C
Tr 22.0 °C
RH 39.1 %
MET 1.20 CLO 1.00
PMV 0.1 PPD 5.1 %
Summary of Contents for HD32.3
Page 4: ...HD32 3 4 V1 4 HD32 3 WBGT PMV PPD indices...
Page 52: ...HD32 3 52 V1 4 12 PRINTING OF THE MEASUREMENT REPORT Norm ISO 7243 Norm ISO 7243...
Page 53: ...HD32 3 53 V1 4 Warm environments Norm ISO 7243...
Page 54: ...HD32 3 54 V1 4 Warm environments Norm ISO 7243 Norm ISO 7243...
Page 55: ...HD32 3 55 V1 4 Warm environments Norm ISO 7243...
Page 56: ...HD32 3 56 V1 4 Warm environments Norm ISO 7243...
Page 57: ...HD32 3 57 V1 4 Warm environments Norm ISO 7243...
Page 58: ...HD32 3 58 V1 4 Moderate environments Norm ISO 7730 Norm ISO 7730...
Page 59: ...HD32 3 59 V1 4 Moderate environments Norm ISO 7730...
Page 60: ...HD32 3 60 V1 4 Moderate environments Norm ISO 7730...
Page 61: ...HD32 3 61 V1 4 Moderate environments Norm ISO 7730...
Page 62: ...HD32 3 62 V1 4 Moderate environments Norm ISO 7730...
Page 63: ...HD32 3 63 V1 4 Moderate environments Norm ISO 7730...
Page 64: ...HD32 3 64 V1 4 Moderate environments Norm ISO 7730...
Page 71: ......