HD32.3
- 45 -
V1.4
Measurement
In the place where you want to measure, you must assembly the tripod and the probes necessary for
the measurement. You set the instrument up and then you start measuring. If the measurement has
to be done in other places, you have to move all the equipment in the new measurement position.
At the end of the measurement or after some times, the acquired data are transferred to the PC for
the elaboration and to allow writing the measurement report/s.
6.1.1 Instruction, accuracy and maintenance of the probes
•
Don’t expose the probes to gas or liquids that could corrode the probe material. After the meas-
urement clean accurately the probes.
•
Don’t fold the connector, using force upwards or downwards.
•
Respect the correct probe polarity.
•
During the insertion of the probe’s connector into the instrument, don’t fold or use force against
the contacts.
•
Don’t fold the probes and don’t distort them or let them fall: they can damage in an irreparable
manner.
•
Use a probe suitable to the measurement type that you want to realize.
•
For a reliable measurement, avoid too speed temperature variations.
•
Some sensors aren’t isolated with reference to the external sheathing, pay attention
not to be in contact with items under tension (up to 48V): it could be dangerous for
the instrument and for the operator who can be electrocute.
•
Avoid measuring in presence of high-frequency sources, microwave or great magnetic fields,
because they will be not so reliable.
•
The instrument is water-proof, don't dip it into the water. If it falls into the water, check that
there isn't any infiltration.
Summary of Contents for HD32.3
Page 4: ...HD32 3 4 V1 4 HD32 3 WBGT PMV PPD indices...
Page 52: ...HD32 3 52 V1 4 12 PRINTING OF THE MEASUREMENT REPORT Norm ISO 7243 Norm ISO 7243...
Page 53: ...HD32 3 53 V1 4 Warm environments Norm ISO 7243...
Page 54: ...HD32 3 54 V1 4 Warm environments Norm ISO 7243 Norm ISO 7243...
Page 55: ...HD32 3 55 V1 4 Warm environments Norm ISO 7243...
Page 56: ...HD32 3 56 V1 4 Warm environments Norm ISO 7243...
Page 57: ...HD32 3 57 V1 4 Warm environments Norm ISO 7243...
Page 58: ...HD32 3 58 V1 4 Moderate environments Norm ISO 7730 Norm ISO 7730...
Page 59: ...HD32 3 59 V1 4 Moderate environments Norm ISO 7730...
Page 60: ...HD32 3 60 V1 4 Moderate environments Norm ISO 7730...
Page 61: ...HD32 3 61 V1 4 Moderate environments Norm ISO 7730...
Page 62: ...HD32 3 62 V1 4 Moderate environments Norm ISO 7730...
Page 63: ...HD32 3 63 V1 4 Moderate environments Norm ISO 7730...
Page 64: ...HD32 3 64 V1 4 Moderate environments Norm ISO 7730...
Page 71: ......