- 21 -
0) Print selected log:
By selecting this item, the page of the log to be printed will be displayed:
1.
Use the arrows
▲▼◄►
to select the log to be printed, and the F1 and F3 function keys to shift
page.
2.
Once a file is selected, the acquisition start date and time and the number of samples contained in
the file (Rec) are displayed at the bottom of the display.
Warning: The files are stored in ascending order. However, if the memory is arranged in
circular mode (please see later at point 3), it could be that a lower number corresponds to a
more recent file
. Each file is only identified by the date and time,
shown on the display
. In the
example above, the file 00 is selected: The recording started at 18:50 on 27 January 2006. The
file contains 6 samples.
3.
Press
ENTER
to print the selected log (or press
ESC
to return to the previous menu, without
printing);
4.
The data transfer message will be displayed a few seconds, then the instrument will go back to
the
Print selected log
page to select another log to be printed;
5.
Repeat the procedure to print the required sessions or press
ESC
to exit this menu;
6.
Press
SETUP
to exit immediately from the main menu.
1) Erase ALL logs
By selecting this item, the “Erase ALL files” message will be displayed:
1.
Press
ENTER
to erase all the files;
2.
Press
ESC
to cancel the operation and return to the previous menu level;
3.
Press
SETUP
to exit immediately from the main menu.
2) Select memory type
By selecting this item, the memory type selection page will be displayed: