HD35
- 106 -
V1.12
Address Type
Input Register description
AP ED RE AL
1016 +
200
x
(N-1) SW
T
EMPERATURE WITH
TC_T sensor of channel N in the set
measurement unit (x10).
1017 +
200
x
(N-1)
B
Alarm for temperature with TC_T sensor of channel N:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
1018 +
200
x
(N-1) SW
T
EMPERATURE
with TC_N sensor of channel N in the set
measurement unit (x10).
1019 +
200
x
(N-1)
B
Alarm for temperature with TC_N sensor of channel N:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
1026 +
200
x
(N-1) SW
T
EMPERATURE
with TC_E sensor of channel N in the set
measurement unit (x10).
1027 +
200
x
(N-1)
B
Alarm for temperature with TC_E sensor of channel N:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
1028 +
200
x
(N-1) SW
Input value in mV of channel N (x10). Only if channel N is
configured as 0…1 V input (HD35EDH).
1029 +
200
x
(N-1)
B
Alarm for channel N if the channel is configured as 0…1 V in-
put (HD35EDH):
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
1030 +
200
x
(N-1) SW
Input value in mV of channel N (x100). Only if channel N is
configured as 0…50 mV input (HD35EDH).
1031 +
200
x
(N-1)
B
Alarm for channel N if the channel is configured as 0…50 mV
input (HD35EDH):
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
1032 +
200
x
(N-1) SW
Input value in mA of channel N (x100). Only if channel N is
configured as 4…20 mA input (HD35EDH).
1033 +
200
x
(N-1)
B
Alarm for channel N if the channel is configured as 4…20 mA
input (HD35EDH):
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
1034 +
200
x
(N-1) SW
Position of potentiometer in % of channel N. Only if channel N
is configured as potentiometric input (HD35EDH).
1035 +
200
x
(N-1)
B
Alarm for channel N if the channel is configured as potentiom-
etric input (HD35EDH):
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
1036 +
200
x
(N-1) SW
Value of quantity associated to channel N if the channel is config-
ured as 0…1 V input (HD35EDH).
1037 +
200
x
(N-1)
B
Alarm for quantity associated to channel N if the channel is con-
figured as 0…1 V input (HD35EDH):
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
1038 +
200
x
(N-1) SW
Value of quantity associated to channel N if the channel is config-
ured as 0…50 mV input (HD35EDH).
1039 +
200
x
(N-1)
B
Alarm for quantity associated to channel N if the channel is con-
figured as 0…50 mV input (HD35EDH):
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
1040 +
200
x
(N-1) SW
Value of quantity associated to channel N if the channel is config-
ured as 4…20 mA input (HD35EDH).
1041 +
200
x
(N-1)
B
Alarm for quantity associated to channel N if the channel is con-
figured as 4…20 mA input (HD35EDH):
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
1042 +
200
x
(N-1) SW
Value of quantity associated to channel N if the channel is config-
ured as potentiometric input (HD35EDH).
1043 +
200
x
(N-1)
B
Alarm for quantity associated to channel N if the channel is con-
figured as potentiometric input (HD35EDH):
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
1044 +
200
x
(N-1) SW
Input value in mV of channel N. Only if channel N is
configured as 0…10 V input (HD35EDWH).
1045 +
200
x
(N-1)
B
Alarm for channel N if the channel is configured as 0…10 V in-
put (HD35EDWH):
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.