HD35
- 113 -
V1.12
Address Type
Holding Register description
AP ED RE AL
Measurement alarm thresholds for configurable inputs
1000 +
200
x
(N-1) SW
Lower alarm threshold for
TEMPERATURE
with 2-wire Pt100
sensor of channel N in the set measurement unit (x10).
1001 +
200
x
(N-1) SW
Higher alarm threshold for temperature with 2-wire Pt100 sen-
sor of channel N in the set measurement unit (x10).
1002 +
200
x
(N-1) SW
Lower alarm threshold for
TEMPERATURE
with 3-wire Pt100
sensor of channel N in the set measurement unit (x10).
1003 +
200
x
(N-1) SW
Higher alarm threshold for temperature with 3-wire Pt100 sen-
sor of channel N in the set measurement unit (x10).
1004 +
200
x
(N-1) SW
Lower alarm threshold for
TEMPERATURE
with 4-wire Pt100
sensor of channel N in the set measurement unit (x10).
1005 +
200
x
(N-1) SW
Higher alarm threshold for temperature with 4-wire Pt100 sen-
sor of channel N in the set measurement unit (x10).
1006 +
200
x
(N-1) SW
Lower alarm threshold for
TEMPERATURE
with 2-wire Pt1000
sensor of channel N in the set measurement unit (x10).
1007 +
200
x
(N-1) SW
Higher alarm threshold for temperature with 2-wire Pt1000
sensor of channel N in the set measurement unit (x10).
1008 +
200
x
(N-1) SW
Lower alarm threshold for
TEMPERATURE
with 3-wire Pt1000
sensor of channel N in the set measurement unit (x10).
1009 +
200
x
(N-1) SW
Higher alarm threshold for temperature with 3-wire Pt1000
sensor of channel N in the set measurement unit (x10).
1010 +
200
x
(N-1) SW
Lower alarm threshold for
TEMPERATURE
with 4-wire Pt1000
sensor of channel N in the set measurement unit (x10).
1011 +
200
x
(N-1) SW
Higher alarm threshold for temperature with 4-wire Pt1000
sensor of channel N in the set measurement unit (x10).
1012 +
200
x
(N-1) SW
Lower alarm threshold for
TEMPERATURE
with TC_K sensor of
channel N in the set measurement unit (x10).
1013 +
200
x
(N-1) SW
Higher alarm threshold for temperature with TC_K sensor of
channel N in the set measurement unit (x10).
1014 +
200
x
(N-1) SW
Lower alarm threshold for
TEMPERATURE
with TC_J sensor of
channel N in the set measurement unit (x10).
1015 +
200
x
(N-1) SW
Higher alarm threshold for temperature with TC_J sensor of
channel N in the set measurement unit (x10).
1016 +
200
x
(N-1) SW
Lower alarm threshold for
TEMPERATURE
with TC_T sensor of
channel N in the set measurement unit (x10).
1017 +
200
x
(N-1) SW
Higher alarm threshold for temperature with TC_T sensor of
channel N in the set measurement unit (x10).
1018 +
200
x
(N-1) SW
Lower alarm threshold for
TEMPERATURE
with TC_N sensor of
channel N in the set measurement unit (x10).
1019 +
200
x
(N-1) SW
Higher alarm threshold for temperature with TC_N sensor of
channel N in the set measurement unit (x10).
1026 +
200
x
(N-1) SW
Lower alarm threshold for
TEMPERATURE
with TC_E sensor of
channel N in the set measurement unit (x10).
1027 +
200
x
(N-1) SW
Higher alarm threshold for temperature with TC_E sensor of
channel N in the set measurement unit (x10).
1028 +
200
x
(N-1) SW
Channel N lower alarm threshold in mV (x10). Only if channel
N is configured as 0…1 V input (HD35EDH).
1029 +
200
x
(N-1) SW
Channel N higher alarm threshold in mV (x10). Only if channel
N is configured as 0…1 V input (HD35EDH).
1030 +
200
x
(N-1) SW
Channel N lower alarm threshold in mV (x100). Only if channel
N is configured as 0…50 mV input (HD35EDH).
1031 +
200
x
(N-1) SW
Channel N higher alarm threshold in mV (x100). Only if chan-
nel N is configured as 0…50 mV input (HD35EDH).
1032 +
200
x
(N-1) SW
Channel N lower alarm threshold in mA (x100). Only if chan-
nel N is configured as 4…20 mA input (HD35EDH).
1033 +
200
x
(N-1) SW
Channel N higher alarm threshold in mA (x100). Only if chan-
nel N is configured as 4…20 mA input (HD35EDH).