Each instrument in the RS485 network is univocally identified by an address ranging within 1
and 247. No more than one transmitter with the same address can be present in the
same network.
Before connecting the base unit to the RS485 network, configure address and Baud Rate (see
chapter " HD35AP… base unit configuration ” of the software online help). The communication
parameters in the PC/PLC must be the same as those set in the base unit.
6.7
HD35APR
CONNECTIONS
RS485 line termination:
If the instrument is at one end of the RS485 line, insert the termination by placing the short
jumper next to the RS485 connector between the “RT” and “CLOSE” indications. Otherwise,
remove the termination by placing the short jumper between the “RT” and “OPEN” indications.
Factory ETHERNET settings:
•
IP address = 192.168.1.235 static
•
Subnet mask = 255.255.255.0
•
Ports = 5100 for proprietary TCP protocol (8 sockets), 502 for Modbus TCP/IP protocol (2
sockets)
The ETHERNET settings can be changed with the HD35AP-S software. It is possible to restore
the factory parameters by placing the short jumper next to the NET RST push-button between
the “2” and “3” indications and then pressing the NET RST push-button. After the reset, replace
the short jumper between the “2” and “1 (NORMAL)” indications.