HD35
- 52 -
V1.12
11) CAL_MENU (calibration) – Only available with administrator password
1)
RH_75%_CAL: relative humidity sensor calibration at 75%RH.
2)
RH_33%_CAL: relative humidity sensor calibration at 33%RH.
3)
HOSE_LEN_m: length (in meters) of the tube connecting the input of the instru-
ment to the pressure measuring point (enter the length of only one of the two
tubes). The item appears only in HD35ED…4r5TV models.
4)
HOSE_DIAM_mm: diameter (in mm) of the tube connecting the input of the in-
strument to the pressure measuring point (default 6 mm). The item appears only in
HD35ED…4r5TV models.
5)
ABS_PRES_mbar: absolute pressure (in mbar) of the air in the connecting tube.
The item appears only in HD35ED…4r5TV models.
6)
DIFF_PRES_0 Pa_CAL: calibration of differential pressure to zero.
7)
LGHT_SENS_PA_LUX: illuminance sensor sensitivity in pA/lux.
8)
UVA_SENS_nA_W/m
2
: UVA irradiance sensor sensitivity in nA/Wm
-2
.
9)
CAL_TYPE: choice between user calibration (USER) or factory calibration (FACT).
10)
EXIT: returns to the main menu.
Note: according to the data logger model, some items could not be available if not sig-
nificant for that particular model.
12) EXIT
Returns to measurement mode.