CALIBRATION WITH THE CYLINDER
To calibrate the CO
2
sensor with the aid of the cylinder, unscrew the probe filter, screw the
HD31.B3A adapter and connect the cylinder; adjust the bottle flow meter to get a constant
flow between 0.3 and 0.5 l/min.