HD37AB17D/HD37B17D
- 11 -
V1.5
Key
Response
Description
A00ZG2
&2008/12/10 08.30.00 ;
2008/12/10 08.50.00 |
Date and hour of factory calibra-
tion
Date and hour of user calibration
A00ZG3
&12345678|
Instrument serial number
A00ZG4
&USERCODE|
User code (max. 32 characters)
A00ZK1 &|
Start
Logging
A00ZK0 &|
Stop
Logging
A00ZKT &|
It enables the compensation in
temperature
A00ZKN
It disables the compensation in
temperature
A00ZKRxxxx yyyy
It executes the memory Dump from
the xxxx page to the yyyy page
(pages: from 0000 to 1999)
A00ZKS
It prints the list of the memo-
rized sections
A00ZKE &|
It resets the number of the sec-
tions and of the records
A00ZKZ &|
It resets the Flag of Automatic
Start
A00ZRL &00000|
Length in seconds of the pro-
grammed Log
A00ZRM
&06800|
Next record number in logging
A00ZRP
&01013|
Set atmospheric pressure
A00ZRS
&00 ; 00 ; FF ; 00 |
Options status:
FIRST BYTE:
temperature correction:
SECOND BYTE:
user correction: 00=off
THIRD BYTE:
Status Log: FF=off 00=start
01=Log 02=end
FOURTH BYTE:
Deferred Log: 00=off
A00ZRT &0000000090| Sections
number
A00ZYU USER
UN_locked|
It enables the user parameters
change
A00ZWUxx…x &|
It writes till 32 characters in
the user code (it’s accepted
only if it is preceded by
A00ZYU)
A00ZWAxx &|
It sets the average per package
on xx samples (from 3 sec to 60
sec)
A00ZWPxxxx
&|0ZWP1013
Setup of atmospheric pressure
Summary of Contents for HD37AB17D
Page 19: ......