HD50
- 30 -
V1.4
Address Type
Input Register description
51 B
Alarm for temperature with NTC10K sensor of channel 3: 0=OFF,
1=lower threshold alarm, 2=higher threshold alarm.
64 SW
T
EMPERATURE
with
Pt100 sensor
of HP3517E… probe in the set measure-
ment unit (x10).
65 B
Alarm for temperature with Pt100 sensor
of HP3517E… probe: 0=OFF,
1=lower threshold alarm, 2=higher threshold alarm.
80 SW
I
LLUMINANCE
in lux (high range, models HD50...I2...).
81 B
Illuminance (high range, models HD50...I2...) alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
Measured values and status of measurement alarms for configurable inputs
(HD50GH)
1000 +
200 x (
N
-1) SW
T
EMPERATURE
with 2-wire
Pt100 sensor
of channel
N
in the set measure-
ment unit (x10).
1001 +
200 x (
N
-1)
B
Alarm for temperature with 2-wire Pt100 sensor
of channel
N
: 0=OFF,
1=lower threshold alarm, 2=higher threshold alarm.
1002 +
200 x (
N
-1) SW
T
EMPERATURE
with 3-wire
Pt100 sensor
of channel
N
in the set measure-
ment unit (x10).
1003 +
200 x (
N
-1)
B
Alarm for temperature with 3-wire Pt100 sensor
of channel
N
: 0=OFF,
1=lower threshold alarm, 2=higher threshold alarm.
1004 +
200 x (
N
-1) SW
T
EMPERATURE
with 4-wire
Pt100 sensor
of channel
N
in the set measure-
ment unit (x10).
1005 +
200 x (
N
-1)
B
Alarm for temperature with 4-wire Pt100 sensor
of channel
N
: 0=OFF,
1=lower threshold alarm, 2=higher threshold alarm.
1006 +
200 x (
N
-1) SW
T
EMPERATURE
with 2-wire
Pt1000 sensor
of channel
N
in the set meas-
urement unit (x10).
1007 +
200 x (
N
-1)
B
Alarm for temperature with 2-wire Pt1000 sensor
of channel
N
: 0=OFF,
1=lower threshold alarm, 2=higher threshold alarm.
1008 +
200 x (
N
-1) SW
T
EMPERATURE
with 3-wire
Pt1000 sensor
of channel
N
in the set meas-
urement unit (x10).
1009 +
200 x (
N
-1)
B
Alarm for temperature with 3-wire Pt1000 sensor
of channel
N
: 0=OFF,
1=lower threshold alarm, 2=higher threshold alarm.
1010 +
200 x (
N
-1) SW
T
EMPERATURE
with 4-wire
Pt1000 sensor
of channel
N
in the set meas-
urement unit (x10).
1011 +
200 x (
N
-1)
B
Alarm for temperature with 4-wire Pt1000 sensor
of channel
N
: 0=OFF,
1=lower threshold alarm, 2=higher threshold alarm.
1012 +
200 x (
N
-1) SW
T
EMPERATURE
with
TC_K
sensor of channel
N
in the set measurement
unit (x10).
1013 +
200 x (
N
-1)
B
Alarm for temperature with TC_K sensor
of channel
N
:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
1014 +
200 x (
N
-1) SW
T
EMPERATURE
with
TC_J
sensor
of channel
N
in the set measurement unit
(x10).
1015 +
200 x (
N
-1)
B
Alarm for temperature with TC_J sensor
of channel
N
:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
1016 +
200 x (
N
-1) SW
T
EMPERATURE
WITH
TC_T
sensor
of channel
N
in the set measurement unit
(x10).
1017 +
200 x (
N
-1)
B
Alarm for temperature with TC_T sensor
of channel
N
:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
1018 +
200 x (
N
-1)
SW
T
EMPERATURE
with
TC_N
sensor
of channel
N
in the set measurement unit
(x10).
1019 +
200 x (
N
-1)
B
Alarm for temperature with TC_N sensor
of channel
N
:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
1026 +
200 x (
N
-1)
SW
T
EMPERATURE
with
TC_E
sensor
of channel
N
in the set measurement unit
(x10).
Summary of Contents for HD50 14b7P TC
Page 52: ...HD50 52 V1 4 NOTES ...
Page 53: ...HD50 53 V1 4 NOTE ...
Page 54: ...HD50 54 V1 4 NOTE ...
Page 55: ......