EN-14
Result Reading (for β-Ketone test)
Message
What it Means
< 0.1 mmol/L
0.1 to 8.0 mmol/L
> 8.0 mmol/L
Error Message
Error Message Cause
What To Do
The batteries cannot
provide enough power
for a test.
Replace the batteries
immediately.
Strip has been used.
Repeat the test with a
new strip.
Problem in operation.
Review the instructions
and repeat the test with
a new strip. If problem
persists, contact the local
customer service for
assistance.
You may have removed
the strip after applying
blood, or insufficient
blood volume.
Review the instructions
and repeat test with a
new test strip.
HCT>75%
Wrong code strip
is inserted or other
coding errors.
Make sure to insert the
right code strip for the
meter. Check if the code
numbers on the code strip
and the strip vial label are
the same.
Expired code strip.
Repeat the test with a
new code strip.
Ambient temperature
is below the system’s
operation range.
System operational range
is 10°C to 40°C (50°F to
104°F). Repeat the test
after the device and test
strip have reached the
above temperature.
Ambient temperature
is above the system’s
operation range.