61
17.
Insert the probe in the container, paying attention that the cap with the probe is in the base. The
measuring chamber has to be perfectly closed, otherwise it won’t reach saturation.
18.
Wait 30 minutes, at least and press OK to confirm.
19.
In the next window, enter the humidity value actually applied and press OK.
20.
Take out the probe from the container with the 33%RH solution, then close it with its lid.
21.
Unscrew the ring with the cap and replace it with the sensor protection. Calibration is finished.