HD32.1
- 4 -
V2.1
1.
GENERAL
CHARACTERISTICS
The
HD32.1
Thermal Microclimate has been designed for microclimate analysis in the workplace;
the instrument is used to detect the necessary parameters to establish if a certain workspace is
suitable to perform certain activities.
The instrument can manage
three operating programs
that can be loaded by the user, according to
the measurements program and the quantities being detected.
The instrument has eight inputs for probes with SICRAM module: The probes are fitted with an
electronic circuit that communicates with the instrument. The calibration settings are memorized
inside.
All SICRAM probes, except the vane probe, can be plugged into any input: They will be
automatically detected when you turn the instrument on.
NOTE: The vane probe, complete with SICRAM module, for wind speed measurement can be
exclusively connected to input 8.
The instrument comes with a barometric pressure sensor. The barometric pressure is displayed only
by the Microclimate Analysis operating program.
The machine can be programmed to perform the logging (capture) of a measurement session. You
can specifically set a sampling interval per each session.
Furthermore, the auto-start function can be used to activate the setting of the measurements’ initial
date and time, and the automatic start and end of the data logging session.
NOTE: The set capture interval is valid for all the probes connected to the machine.
Other operating programs user selectable/settable common parameters are:
•
The units of measurement for the displayed temperature quantities: °C, °F, °K.
•
The system date and time
•
The display of the
maximum, minimum, and average
statistic parameters and their
deletion.
•
The data transfer speed via the serial RS232 port.
•
The setting and enabling/disabling of the keyboard protection password.
The operating programs are:
•
prog. A: HD32.1 Microclimate Analysis
•
prog. B: HD32.1
Discomfort Analysis
•
prog. C: HD32.1
Physical Quantities
The
operating
program
A: Microclimate Analysis,
HD32.1, can simultaneously detect the
following quantities:
•
Globe thermometer temperature
•
Natural ventilation wet bulb temperature
•
Environment temperature
•
Atmospheric pressure
•
Relative humidity
•
Wind speed
Summary of Contents for HD32.1
Page 6: ...HD32 1 6 V2 1 Thermal Microclimate HD32 1...
Page 78: ...HD32 1 78 V2 1 MIN 350 MAX 1020 120 AP471 S4...
Page 106: ...HD32 1 106 V2 1 Moderate Environment Norm ISO 7730...
Page 107: ...HD32 1 107 V2 1 Moderate Environment Norm ISO 7730...
Page 108: ...HD32 1 108 V2 1 Moderate Environment Norm ISO 7730...
Page 109: ...HD32 1 109 V2 1 Moderate Environment Norm ISO 7730...
Page 110: ...HD32 1 110 V2 1 Moderate Environment Norm ISO 7730...
Page 111: ...HD32 1 111 V2 1 Moderate Environment Norm ISO 7730...
Page 112: ...HD32 1 112 V2 1 12 2 WARM ENVIRONMENT Norm ISO 7243 Norm ISO 7243...
Page 113: ...HD32 1 113 V2 1 Warm Environment Norm ISO 7243...
Page 114: ...HD32 1 114 V2 1 Warm Environment Norm ISO 7243 Norm ISO 7243...
Page 115: ...HD32 1 115 V2 1 Warm Environment Norm ISO 7243...
Page 116: ...HD32 1 116 V2 1 Warm Environment Norm ISO 7243...
Page 117: ...HD32 1 117 V2 1 Warm Environment Norm ISO 7243...
Page 118: ...HD32 1 118 V2 1 12 3 DISCOMFORT ANALYSIS Norm ISO 7730 Norm ISO 7730...
Page 119: ...HD32 1 119 V2 1 Discomfort Analysis Norm ISO 7730...
Page 120: ...HD32 1 120 V2 1 Discomfort Analysis Norm ISO 7730...
Page 121: ...HD32 1 121 V2 1 Discomfort Analysis Norm ISO 7730...
Page 122: ...HD32 1 122 V2 1 Discomfort Analysis Norm ISO 7730...
Page 123: ...HD32 1 123 V2 1 Discomfort Analysis Norm ISO 7730...
Page 124: ...HD32 1 124 V2 1 Discomfort Analysis Norm ISO 7730...
Page 125: ...HD32 1 125 V2 1 12 4 HOT ENVIRONMENT Norm ISO 7933 Norm ISO 7933...
Page 126: ...HD32 1 126 V2 1 Hot Environment Norm ISO 7933...
Page 127: ...HD32 1 127 V2 1 Hot Environment Norm ISO 7933...
Page 128: ...HD32 1 128 V2 1 Hot Environment Norm ISO 7933...
Page 129: ...HD32 1 129 V2 1 Hot Environment Norm ISO 7933...
Page 130: ...HD32 1 130 V2 1 Hot Environment Norm ISO 7933...
Page 131: ...HD32 1 131 V2 1 Hot Environment Norm ISO 7933...
Page 132: ...HD32 1 132 V2 1 12 5 COLD ENVIRONMENT Norm ISO 11079 Norm ISO 11079...
Page 133: ...HD32 1 133 V2 1 Cold Environment Norm ISO 11079...
Page 134: ...HD32 1 134 V2 1 Cold Environment Norm ISO 11079...
Page 135: ...HD32 1 135 V2 1 Cold Environment Norm ISO 11079...
Page 136: ...HD32 1 136 V2 1 Cold Environment Norm ISO 11079...
Page 137: ...HD32 1 137 V2 1 Cold Environment Norm ISO 11079...
Page 138: ...HD32 1 138 V2 1 Cold Environment Norm ISO 11079...
Page 139: ...HD32 1 139 V2 1 Cold Environment Norm ISO 11079...
Page 140: ...HD32 1 140 V2 1 12 6 PHYSICAL QUANTITIES...
Page 153: ...HD32 1 153 V2 1 NOTES...
Page 154: ...HD32 1 154 V2 1 NOTES...
Page 155: ......