The
operating
program
A: Microclimate Analysis
displays also:
•
The local turbulence intensity
Tu
, for DR (Draught Rating) calculation.
•
The
WBGT
index (Wet Bulb Glob Temperature) with or without solar radiation.
•
The
WCI
index (Wind Chill Index)
•
The average radiation temperature
t
r
.
The
operating
program
B: Discomfort Analysis,
HD32.1, can simultaneously detect the
following quantities:
•
Air temperature detected at head height (1.7 m for a standing person; 1.1 m for a seated
person).
•
Air temperature detected at abdomen height (1.1 m for a standing person; 0.6 m for a seated
person).
•
Air temperature detected at ankle height (0.1 m).
•
Temperature at floor level.
•
Temperature of the net radiometer.
•
Net radiation.
•
Radiant asymmetry temperature.
The
operating
program
B:
Discomfort Analysis
is used to calculate the local discomfort indexes
due to vertical temperature gradients or radiant asymmetry temperature.
The
operating
program
C: Physical Quantities,
HD32.1, can simultaneously detect the following
physical quantities:
•
Temperature
•
Relative humidity
•
Illuminance, luminance, PAR, irradiance
•
Wind speed
•
carbon monoxide concentration CO.
•
carbon dioxide concentration CO
2
By using the
operating
program
C: Physical Quantities,
HD32.1
can simultaneously manage
up to six different probes
complete with SICRAM module: one or two light probes and a probe for
each of the other physical quantities.
If two or more probes of the same physical quantity are
connected (light probes excluded), the instrument recognizes only one probe
. If, for example,
you insert two Pt100 temperature probes complete with SICRAM module to inputs 1 and 2, the
probe connected to input 1 is immediately recognized while the probe connected to input 2 is
ignored.
The probe is detected during turn on
, therefore if a probe is changed, it is necessary to
turn the machine off and on. During turn on the instrument performs a scan from input 1 to input 8.
The probes arrangement in relation to the inputs is arbitrary, except for the vane probes.
The vane probe, complete with SICRAM module, for wind speed measurement can be
exclusively connected to input 8.
Summary of Contents for HD32.1
Page 6: ...HD32 1 6 V2 1 Thermal Microclimate HD32 1...
Page 78: ...HD32 1 78 V2 1 MIN 350 MAX 1020 120 AP471 S4...
Page 106: ...HD32 1 106 V2 1 Moderate Environment Norm ISO 7730...
Page 107: ...HD32 1 107 V2 1 Moderate Environment Norm ISO 7730...
Page 108: ...HD32 1 108 V2 1 Moderate Environment Norm ISO 7730...
Page 109: ...HD32 1 109 V2 1 Moderate Environment Norm ISO 7730...
Page 110: ...HD32 1 110 V2 1 Moderate Environment Norm ISO 7730...
Page 111: ...HD32 1 111 V2 1 Moderate Environment Norm ISO 7730...
Page 112: ...HD32 1 112 V2 1 12 2 WARM ENVIRONMENT Norm ISO 7243 Norm ISO 7243...
Page 113: ...HD32 1 113 V2 1 Warm Environment Norm ISO 7243...
Page 114: ...HD32 1 114 V2 1 Warm Environment Norm ISO 7243 Norm ISO 7243...
Page 115: ...HD32 1 115 V2 1 Warm Environment Norm ISO 7243...
Page 116: ...HD32 1 116 V2 1 Warm Environment Norm ISO 7243...
Page 117: ...HD32 1 117 V2 1 Warm Environment Norm ISO 7243...
Page 118: ...HD32 1 118 V2 1 12 3 DISCOMFORT ANALYSIS Norm ISO 7730 Norm ISO 7730...
Page 119: ...HD32 1 119 V2 1 Discomfort Analysis Norm ISO 7730...
Page 120: ...HD32 1 120 V2 1 Discomfort Analysis Norm ISO 7730...
Page 121: ...HD32 1 121 V2 1 Discomfort Analysis Norm ISO 7730...
Page 122: ...HD32 1 122 V2 1 Discomfort Analysis Norm ISO 7730...
Page 123: ...HD32 1 123 V2 1 Discomfort Analysis Norm ISO 7730...
Page 124: ...HD32 1 124 V2 1 Discomfort Analysis Norm ISO 7730...
Page 125: ...HD32 1 125 V2 1 12 4 HOT ENVIRONMENT Norm ISO 7933 Norm ISO 7933...
Page 126: ...HD32 1 126 V2 1 Hot Environment Norm ISO 7933...
Page 127: ...HD32 1 127 V2 1 Hot Environment Norm ISO 7933...
Page 128: ...HD32 1 128 V2 1 Hot Environment Norm ISO 7933...
Page 129: ...HD32 1 129 V2 1 Hot Environment Norm ISO 7933...
Page 130: ...HD32 1 130 V2 1 Hot Environment Norm ISO 7933...
Page 131: ...HD32 1 131 V2 1 Hot Environment Norm ISO 7933...
Page 132: ...HD32 1 132 V2 1 12 5 COLD ENVIRONMENT Norm ISO 11079 Norm ISO 11079...
Page 133: ...HD32 1 133 V2 1 Cold Environment Norm ISO 11079...
Page 134: ...HD32 1 134 V2 1 Cold Environment Norm ISO 11079...
Page 135: ...HD32 1 135 V2 1 Cold Environment Norm ISO 11079...
Page 136: ...HD32 1 136 V2 1 Cold Environment Norm ISO 11079...
Page 137: ...HD32 1 137 V2 1 Cold Environment Norm ISO 11079...
Page 138: ...HD32 1 138 V2 1 Cold Environment Norm ISO 11079...
Page 139: ...HD32 1 139 V2 1 Cold Environment Norm ISO 11079...
Page 140: ...HD32 1 140 V2 1 12 6 PHYSICAL QUANTITIES...
Page 153: ...HD32 1 153 V2 1 NOTES...
Page 154: ...HD32 1 154 V2 1 NOTES...
Page 155: ......