HD52.3D
- 4 -
V1.11
2
TECHNICAL SPECIFICATIONS
Wind speed
Used sensor
Ultrasounds
Measuring range
0…60 m/s
Resolution 0.01
m/s
Accuracy ±
0.2 m/s or ±
2%, the greatest (0…35
m/s), ± 3% (>
35
m/s)
Wind direction
Used sensor
Ultrasounds
Measuring range
0…359.9°
Resolution 0.1°
Accuracy
± 2° RMSE from 1.0 m/s
Compass
Used sensor
Magnetic
Measuring range
0…360°
Resolution 0.1°
Accuracy
± 1°
Air temperature
(
needs option
17
)
Used sensor
Pt100
Measuring range
-40…+60 °C
Resolution
0.1 °C
Accuracy
± 0.15 °C ± 0.1% of measurement
Relative humidity
(
needs option
17
)
Used sensor
Capacitive
Measuring range
0…100%RH
Resolution
0.1%
Accuracy (@ T = 15…35 °C)
±
1.5%RH (0…90%RH),
±
2%RH (remaining range)
Accuracy (@ T = -40…+60 °C)
±
(1.5 + 1.5% of measurement )%RH
Barometric Pressure
(
needs option
4
)
Used sensor
Piezoresistive
Measuring range
300…1100 hPa
Resolution
0.1 hPa
Accuracy
± 0.5 hPa @ 20
°C
Solar radiation
(
needs option
P
)
Used sensor
Thermopile
Measuring range
0…2000 W/m
2
Resolution
1 W/m
2
Accuracy
2
nd
Class Pyranometer
General features
Power Supply
10…30 Vdc
Power consumption
26
mA @ 12
Vdc without heater, 6
W with heater
Serial outputs
RS232, RS485 (¼ Unit Load), RS422 and SDI-12
Communication protocols
NMEA, MODBUS-RTU, SDI-12, proprietary RS232 and RS485
Analog outputs
2 analog outputs, for wind speed and direction. Output at
choice among 4…20
mA (standard), 0…1
V, 0…5
V and
0…10
V (
option 0…10
V needs 15…30
Vdc power supply
)
Electrical connection
19-pole M23 male connector
Operating temperature
-40…+60 °C
Protection degree
IP 64
Dimensions
H=179
mm,
∅
=150
mm (HD52.3D, HD52.3D4)
H=200
mm,
∅
=150
mm (HD52.3DP, HD52.3DP4)
H=336
mm,
∅
=150
mm (HD52.3D17, HD52.3D147)
H=357
mm,
∅
=150
mm (HD52.3DP17, HD52.3DP147)
Weight
About 1 kg (complete version HD52.3DP147)