background image

 

                       

             

 

 

 

1.Purpose 

 

 

 

 

This report is used to calculate the relevant test results of 22362 temperature monitor.   

2. General Specification   

Item

 

B9D 

Usage Type 

Single-use 

Sensor 

Internal digital temperature sensor 

Memory Capacity 

28800 readings 

Measurement Range 

-20°C ~ 70°C        (-4°F ~ 158°F) 

Accuracy 

±0.5°C/0.9°F ( -10°C-45°C )

±1°C (Other) 

Resolution 

0.1

 

Cellular Communications 

2G: GSM (850, 900, 1800, 1900 MHz) 

SIM Card 

Internal Embedded (Global Data Roaming) 

Shelf Life 

1 Year 

Connector 

USB 2.0 

 

3. LED instructions. 

NO 

Test item 

Test method 

          Judge standard 

Result 

LED 

in 

Idle 

Mode 

Press start button for 1 
second when the power 
of the logger is normal 

LED toggle red/blue twice 

 

Pass 

Press start button for 1 
second    when the power 
of the logger is low 

No LED light under any 
condition 

 

Pass 

Press and hold Flight 
button for 5 seconds 

No LED light and logger 
can not enter Flight Mode

 

Pass 

Press and hold start 
button for 5 seconds then 
logger from Idle mode to 
Recording mode 

The status LED flash Blue 
every 10 seconds for 30 
minutes to indicate it has 
started properly After 30 

 
 

Pass 

Reviews: