3
1.Purpose
This report is used to calculate the relevant test results of 22362 temperature monitor.
2. General Specification
Item
B9D
Usage Type
Single-use
Sensor
Internal digital temperature sensor
Memory Capacity
28800 readings
Measurement Range
-20°C ~ 70°C (-4°F ~ 158°F)
Accuracy
±0.5°C/0.9°F ( -10°C-45°C )
,
±1°C (Other)
Resolution
0.1
℃
Cellular Communications
2G: GSM (850, 900, 1800, 1900 MHz)
SIM Card
Internal Embedded (Global Data Roaming)
Shelf Life
1 Year
Connector
USB 2.0
3. LED instructions.
NO
Test item
Test method
Judge standard
Result
1
LED
in
Idle
Mode
Press start button for 1
second when the power
of the logger is normal
LED toggle red/blue twice
Pass
2
Press start button for 1
second when the power
of the logger is low
No LED light under any
condition
Pass
3
Press and hold Flight
button for 5 seconds
No LED light and logger
can not enter Flight Mode
Pass
4
Press and hold start
button for 5 seconds then
logger from Idle mode to
Recording mode
The status LED flash Blue
every 10 seconds for 30
minutes to indicate it has
started properly After 30
Pass