830B 

DMM OPERATOR’S MANUAL

 

 

1. 

Overview

   

This multimeter is characterized at slim size, portable, stable performance . 
Using 3½  digits LCD monitor with character 13MM high, they offer  clear 
readings.    With  overall  circuitry  design  centering  on  large-scale  IC  A/D 
converters in conjunction and over-load protection circuit, the meters give 
excellent performance and exquisite making as a handy utility instrument. 
The  meters  can  be  used  to  measure  DC  &  AC  voltage,  DC  current, 
resistance, positive diode voltage fall, hFE parameters for transistor and 
Continuity.

 

2. Safety Information

 

2-1 The meters are designed according to IEC-1010 concerning electronic 

measuring instruments with an over-voltage category (CAT 

) and 

pollution 2. 

2-2    Follow all safety and operating instructions to ensure that the meter 

is used safely and is kept in good operating condition. 

2-3    safety symbols: 

  important safety information, refer to the operating manual. 
  Dangerous voltage may be presence. 

 

Double insulation (protection Class II) 

3. Special Cautions for Operation 

 

3-1 The meters can be safe only according to standard procedures when 

used  in  conjunctions  with  the  supplied  test  leads.  To  replace 
damaged  test  leads  with  only  the  same  model  or  same  electric 
specifications. 

3-2    To avid risk of electric shock, do not use the meters before the cover 

is in place. 

3-3    The range switch should be right position for the testing. 
3-4    To  avoid  electric  shock  and  damaging  the  instruments,  the  input 

signals are forbidden to exceed the specified limits. 

3-5    When measuring TV set or switched power, attention should be paid 

to the possible pulses that may bring destruction to the circuit. 

3-6    Range switch position is forbidden to be changed at random during 

measurement. 

3-7    Take  caution  against  shock  in  the  course  ot  measuring  voltage 

higher than DC 60V & AC 30V. 

3-8    Protection fuse should be replaced only with same type and same 

specification. 

4. GENERAL SPECIFICATIONS   

4-1 Max Voltage between input terminal and Earth Ground: CAT II 600V   
4-2 Over-range Indication: display “1” for the significant digit. 
4-3 Automatic display of negative polarity “_” 
4-4 Low Battery Indication: „ ‟ displayed 
4-5 Max LCD display: 1999 (31/2 digits) 
4-6 Fuse protection: F-200mA/250V (Ø 5x20MM) 
4-7 Power Supply: 9V battery, 6F22 or NEDA 1604 
4-8 Operating Temp.: 0

  to 40

  (relative humidity <85%) 

4-9 Storage Temp.:-10

  to 50

  ((relative humidity <85%) 

4-10 Guaranteed precision Temp.: 23

±

 

                          (relative humidity <85%) 

4-11 Dimension: 69x126x24 
4-12 Weight: approx. 120g (including battery) 

5. Testing Specifications 

 

Accuracy is specified for a period of year after calibration and at 18

  to 

28

  (64

  to 82

) with relative humidity to 75%. 

5-1    DC Voltage

 

range 

resolution 

accuracy 

200mV 

0.1mV 

±

(0.8% of rdg +10 digits ) 

2000mV 

1mV 

±

(0.8% of rdg + 10digits) 

20V 

10mV 

±

(0.8% of rdg +10 digits) 

200V 

100mV 

±

(0.8% of rdg + 10 digits) 

1000V 

1V 

±

(1.0% of rdg +15 digits) 

-- Input Impendence: 1M

Ω

 

-- Overload protection: 250V for 200mV range, effective DC or AC 750V for 

other ranges 

5-2    AC Voltage 

range 

resolution 

accuracy 

200V 

100mV 

±

(1.2% of rdg + 20 digits) 

750V 

1V 

±

(2.0% of rdg + 20 digits) 

Frequency Range: 40 to 400Hz 

40 to 100Hz for range 750V 

-- Response: average, calibrated in rms of sine wave

 

5-3    DC Current 

range 

Resolution 

accuracy 

200uA 

0.1

μ

±

(1.0% of rdg + 15 digits) 

2000uA 

1

μ

20mA 

10

μ

200mA 

100

μ

±

(1.5% of rdg + 20 digits) 

10A 

10mA 

±

(3.0% of rdg + 20 digits) 

-- Overload protection: F 250mA/250V fuse 

Note: 10A range: not fused.

 

10A up to 10 seconds 

5- 4    Resistance 

range 

Resolution 

accuracy 

200

Ω

 

0.1

Ω

 

±

(1.5% of rdg +25 digits) 

2000

Ω

 

1

Ω

 

±

(0.8% of rdg + 20digits) 

20K

Ω

 

10

Ω

 

±

(0.8% of rdg + 20 digits) 

200K

Ω

 

100

Ω

 

±

(0.8% of rdg +20 digits) 

2000K

Ω

 

1K

Ω

 

±

(1.8% of rdg + 20 digits) 

-- over-load protection: 250V effective value 

5-5 

 

Transistor hFE Test   

range 

Test Range 

Test Current / voltage 

NPN & PNP 

0-1000 

Ib=10uA / Vce=3V 

5-6    Diode Test 

range 

resolution 

Function 

 

1mV 

Display: read approximate 

forward voltage of diode 

-- over-load Protection: 250V effective value 

-- forward DC current: approximate 1mA 

-- Reversed DC voltage: approximate 3.0V 

6. OPERATING INSTRUCTIONS   
6-1 Attention before operation   

6-1-1 Check 7V battery. If the battery voltage is less than 7V, display will 

show  “ ”,  the  battery  should  be  replaced  at  this  time  to  ensure 
measuring precision. 

6-1-2 Pay attention to the “

” besides the input jack which shows that 

the input voltage or current should be within the specified value. 

6-1-3  The  range  switch  should  be  positioned  to  desired  range  for 

measurement before operation. 

6-2 Measuring DC Voltage 

6-2-1 Connect the black test lead to COM jack and the red to V

Ω

mA jack. 

6-2-2 Set the rotary switch at the desired DCV

 

range position. 

6-2-3 Connect test leads across the source or load under 

Reviews: