Pro-Select Platinum®
Operation and Instruction Manual
14
Solution bottles
Warning
PSP® bottles are not autoclavable.
Bottles can be wiped clean with a pH neutral surface disinfectant by following
disinfectant manufacturer’s recommendations.
Handpiece cable
After use flush water through cable, wipe cable with a clean cloth dampened
with a surface disinfectant
or
• Roll into autoclave bag for sterilization, avoid crimping
• Follow the autoclave manufacturer’s recommended sterilization procedures
• Cables routinely autoclaved will require more frequent replacement
Handpieces and LED lights
1. Clean outer surface of handpiece and LED light by wiping with a soft cloth
dampened with pH neutral surface disinfection solution
Attention
• Do not use disinfection solution containing phenols.
• Do not spray cleaning solution directly on the handpiece.
2. Remove LED light from handpiece, place handpiece and LED light in a
cassette or sterilization bag and seal
3. Steam sterilize in accordance with autoclave manufacturer’s instructions
Scaler tips
Clean, disinfect, inspect and steam sterilize, per steam autoclave manufacturer’s instructions, before
each use
Tip change kit
1. Tip wrench and tip grip have been designed to be sterilized in a steam autoclave
2. Place tip wrench and tip grip in an autoclave bag and sterilize by following
autoclave manufacturer’s instructions.
Any method of sterilization not specifically recommended in this Operation and Instruction
Manual should be approved by DenMat prior to its use, in order to prevent voiding of warranty
!