10. Absorbance OD 600 App
This application is generally used to determine the optical density of microbial
cell cultures at 600 nm. Conversions to cells per mL are based upon a user-
defined factor.
Quick Protocol
1. Launch the app from the
Home
screen.
2. Establish a
Blank
using the appropriate buffer.
3. Enter a sample name.
4. Enter a cell conversion factor (optional).
5. Measure a fresh aliquot of sample using the
Measure
button.
Notes:
•
All cell type specific conversion factors are user-derived.
•
Ensure culture is well mixed prior to sampling.
•
Use fresh 1 uL aliquots for each microvolume measuremen
t.
•Although microvolume OD 600 measurements can be made, the cuvette 10 mm mode
is recommended for this app. This option is available on the DS-11 FX +.
43
OD 600
!