Basic Fluorometer
Excitation Selection and Emission Results
•
Select a fluorophore from the drop-down menu in the left panel. The excitation
source is automatically determined for preconfigured fluorophores.
•
The selected LED is the
only
excitation source used during a measurement.
However, the emission RFU will be collected for each longer wavelength
emission filter set.
• For example, if the blue LED is selected, the the emission results for the
blue, green and red channels will be reported. If the green LED is
selected, only the green and red channel results will be reported.
•
New fluorophores may be entered using the List Add/Edit button.
• The Add button will pop-up a window that includes a LED Wizard. Tapping
the Wizard button will result in the sample being excited by each LED and
the RFU measured by the corresponding detector. The LED that resulted
in the largest RFU at its detector will automatically be chosen as the LED
for the method.