User Manual
6
EverLight®
(Post Mount)
Section II Installation
Post Mount for Magellan CMU
3
. Route the supplied power cord through the
bottom of the console and into the front of
the chair.
(P-clamps provided in bag of supplies.)
5
. Roll up any excess wires and neatly place
the power supply and excess wires inside the
console making sure no wires interfere with
closure of the console covers.
6
. Replace the side covers on the console.
7
.
Reconnect
the chair power and test the light
switch as described in Section III Operation.
Light Arm Assembly
1
. Place the chair in its lowest position.
2
.
Disconnect
the chair power.
3
. Apply a light coat of lubricant to the spindle.
4
. Slide the bushing
(lipped end first)
over the light
cord and onto the spindle.
5
. Feed the light cord through the center of the
light pole, pulling the cord out of the bottom
of the pole.
Power Supply Box
1
. Plug the supplied power cord into the power
supply.
(See photo below.)
2
. Plug the light harness from the light arm into
the power supply.
3
. Route the supplied power cord harness
through the front of the seat casting and plug
it into the chair harness.
(See second photo on this
page.)
4
. Roll up any excess wire and neatly place them
inside the seat casting making sure no wires
will interfere with moving parts.
(Wire ties
provided in bag of supplies.)
5
. Using the supplied Velcro, adhere the power
supply to the magellan manifold cover.
4
. Plug the supplied power cord into the chair
harness located inside the chair.
P-Clamp
Console
Supplied
Power Cord
Chair Harness
Supplied
Power Cord
NOTE:
Refer to the Magellan Unit Installation
Manual for installation of the light pole.
NOTE:
If the chair's seat upholstery was installed,
remove it according to the respective chair's manual.
Magellan Manifold
Cover
Power
Supply
6
. Reconnect the chair power and test the light
switch as described in Section III Operation.
NOTE:
Replace the seat upholstery if it was removed.
Power
Cord