CAUTION

CAUTION

4

Preoperational checks

•  The disposable prophy angle and handpiece sleeve are disposable items intended 

for a single patient use only. Do not use the same prophy angle or handpiece sleeve 
for more than one patient. Handle the used disposable prophy angle and handpiece 
sleeve as contaminated patient care items.  Dispose of these items according to 
your specific state and local medical waste regulations.

•  Make sure to install a new single-use handpiece sleeve before every use to prevent 

contamination of the handpiece motor.

•  Do not insert the plug into any outlet other than the one for commercial power 

supply.

•  Turn off the handpiece motor before charging.
•  The standard charging time is Approx. 90 minutes. This may vary depending on 

the battery condition, age, temperature and other factors. With older batteries the 
charging time and operating time may become significantly shorter.

•  Only use the supplied AC adapter.
•  Gently  place  the  handpiece  motor  into  the  stand.  A  strong  impact  may  cause 

handpiece failure.

•  When placing the handpiece motor into the stand pay attention to the insertion 

direction.  Excessive  force  is  not  required  and  it  may  cause  damage  to  the 
handpiece. The handpiece may also be damaged if dropped.

•  Remove the handpiece sleeve from the  handpiece motor before placing it on the 

battery charger.

•  The handpiece battery charging stand is exclusively for the iStar handpiece motor. 

Never use the stand for any other purpose.

•  The battery will get warm during charging, but this is not a failure. Inserting or 

removing the plug at short intervals (approx. 5 minutes) may generate heat and 
interfere with charging. Charge the handpiece for at least 10 minutes.

•  If  not  used  for  a  long  period  charge  the  battery  prior  to  use.  If  not  charged 

periodically the battery will become fully discharged and will not charge properly. If 
this occurs contact your authorized StarDental dealer to purchase a new battery.

Before use operate the product outside the patient's mouth and check that the disposable handpiece 
head rotates properly. If abnormal vibration, noise or overheating is detected during testing or use 
immediately stop use and contact your Authorized StarDental Dealer.

5

Operation

5-1 Charging the Battery

1) Insert the AC adapter plug into a commercial AC power outlet.
2) Insert the handpiece motor into the stand. The battery charge indicator lamp lights up and 

charging starts.

3) When the battery charge indicator lamp turns off charging is completed.

The 

  lamp  lights  up  when  remaining  battery  power  becomes  low.  When  the 

  lamp  is 

illuminated immediately charge the battery.

8

Summary of Contents for istar

Page 1: ...OM EZ721E ...

Page 2: ......

Page 3: ... AC Adapter Plug Head 10 9 Periodic Maintenance 11 10 Overload Protection 11 11 Troubleshooting 12 12 Specifications 12 13 Classification 13 14 Symbols 13 15 Warranty 14 16 Spare Parts 14 17 Disposal 15 18 Electromagnetic Compatibility Information 15 French 17 User and Indications for Use Do not use this product in the vicinity of explosive substances and flammable materials It is the responsibili...

Page 4: ... your Authorized StarDental Dealer for service Using the handpiece under these abnormal conditions may cause injury or damage to property Should battery fluid leak and get into patient s or user s eyes immediately flush eyes thoroughly with clean water and seek medical attention If left untreated this condition may result in blindness Do not charge use or store the product in direct sunlight a veh...

Page 5: ...ospitals and for visiting dental professionals Read this Operation Manual before use to fully understand the product functions Always consider the safety of the patient when operating this product Connect or disconnect the disposable prophy angle only after rotation of the handpiece has completely stopped This may cause injury or cause damage to the connecting portion of the disposable handpiece D...

Page 6: ...e install the AC adapter in such a way that it can be quickly disconnected from the outlet in an emergency Do not place any object within 6 inches of the AC adapter When disconnecting the AC adapter be sure to hold the adapter body Disconnecting by pulling on the cord may damage the cord Refer to the disposable prophy angle operation manual for the specific disposable prophy angle being used for h...

Page 7: ...5 1 Package Contents Set components Product name Quantity 1 Handpiece motor 1 2 Stand 1 3 AC adapter 1 4 Handpiece sleeve 100pcs 1 iStar ...

Page 8: ...otor and pressing it again stops the rotation Alternate operation Alternatively pressing and holding down the ON OFF key rotates the handpiece motor while the button is being pressed Releasing the ON OFF key stops rotation Temporary rotating operation Indicators on operation panel A SPEED lamp The speed is indicated by lamps B Battery charge indicator This indicator lamp lights up during battery c...

Page 9: ...ar Always wear safety glasses and dust mask during use 1 Connecting 1 Turn off the handpiece motor 2 Cover the handpiece with the supplied single use handpiece sleeve Fig 2 3 Connect the disposable prophy angle firmly to the handpiece Check that the connection is secure by running the handpiece outside of the patient s mouth Fig 3 4 2 When operation is complete turn off device and disconnect the d...

Page 10: ...ce may also be damaged if dropped Remove the handpiece sleeve from the handpiece motor before placing it on the battery charger The handpiece battery charging stand is exclusively for the iStar handpiece motor Never use the stand for any other purpose The battery will get warm during charging but this is not a failure Inserting or removing the plug at short intervals approx 5 minutes may generate ...

Page 11: ...attery voltage However the low voltage may not be immediately detected if the product is left under the low voltage condition before turning on the power This is due to the properties of the battery and is not a failure 5 2 Power supply Press and hold down the POWER key to turn on the power The SPEED lamp will light up on the operation panel 5 3 SPEED Setting Speed is adjusted by pressing the SPEE...

Page 12: ... every patient by wiping down with isopropyl alcohol solution Wipe all surfaces including the chuck buttons and bottom charge area as all of these locations may become contaminated during use Compliance with cleaning procedures is critical to patient safety Do not lubricate the handpiece motor Do not use any solvents such as benzene or thinner to clean the handpiece motor This product cannot be st...

Page 13: ...display an error code on the display When an error is displayed turn the power off and on again to check whether the error persists If the error is displayed again follow the solutions recommended in the table below Item Error display Error description Cause Check and remedy During rotation The SPEED lamp flashes Outside the range of working temperatures The use temperature is significantly low Us...

Page 14: ... condensation The temperature of the battery pack is high The battery is not charged when the temperature of the battery pack is approx 40 C 104 F or higher Charge the rechargeable battery within permissible temperatures Handpiece motor stand AC adapter Trouble Cause Remedy The motor does not rotate The handpiece motor has an internal failure Contact your Authorized StarDental Dealer Short circuit...

Page 15: ...n 13 14 Symbols Consult operation instructions Class II equipment Type B applied part Caution Refer to attached instructions Direct Current Alternate Current For indoor use only Efficiency Level TUV Certification Mark Do not reuse Marking on the outside of Equipment or Equipment parts that include RF transmitters or that apply RF electromagnetic energy for diagnosis or treatment This product meets...

Page 16: ...NY WAY OR TO GRANT ANY OTHER WARRANTY Provided conditions in warranty are met warranty Registration Please visit one of the websites below to register your StarDental product This will ensure proper registration and prompt and accurate service should any of your purchases need evaluation or repair during the warranty period For Identafi Oral Cancer Screening System products www identafi net warran...

Page 17: ...The product is intended for use in the electromagnetic environment specified below The customer or the user of the product should assure that it is used in such an environment NOTE Ut is the AC mains voltage prior to application of the test level Immunity test IEC60601 test level Electromagnetic environment guidance Electrostatic discharge ESD IEC61000 4 2 Electrical fast transient burst IEC61000 ...

Page 18: ...e transmitter in watts W according to the transmitter manufacturer and is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Complies with Maximum length Cabl...

Page 19: ......

Page 20: ...C 2015 10 20 ...

Reviews: