Disposal
Dispose of the stand and AC adapter as industrial waste according to local laws and regulations.
Dispose of the handpiece motor as medical waste.
The handpiece motor includes a battery pack (lithium-ion battery). Do not dispose of in a fire or
heat it. Doing so may cause rupture of the battery pack, scattering of the battery fluid, overheating,
smoking, rupture, or explosion.
The battery pack (lithium-ion battery) included in the handpiece motor is a recyclable precious
resource. When disposing of it do not disassemble the battery pack; dispose of the product according
to laws and regulations. Alternatively, return it to your dealer.
18
electromagnetic Compatibility Information
Guidance and manufacturer's declaration - Electromagnetic Emissions
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure that it is used in
such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR11
RF emissions
CISPR11
Harmonic emissions
IEC61000-3-2
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Group 1
Class B
Not Applicable
Not Applicable
The product uses RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
The product is suitable for use in all establishments, including domestic establishments and
those directly connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Guidance and manufacturer's declaration - Electromagnetic Immunity
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure that it is used in
such an environment.
NOTE: ‘Ut’ is the AC mains voltage prior to application of the test level.
Immunity test
IEC60601 test level
Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC61000-4-2
Electrical fast transient/burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC61000-4-11
Power frequency
(50/60Hz)magnetic field
IEC61000-4-8
±(2, 4, 6)kV contact
±(2, 4, 8)kV air
±2kV for power supply lines
±1kV for input/output lines
±1kV line(s) to line(s)
±2kV line(s) to earth
<5% Ut (>95% dip in Ut)
for 0.5 cycles
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (<95% dip in Ut)
for 5 sec
3 A/m
Compliance level
±(2, 4, 6)kV contact
±(2, 4, 8)kV air
±2kV for power supply lines
±1kV for input/output lines
±1kV line(s) to line(s)
±2kV line(s) to earth
<5% Ut (>95% dip in Ut)
for 0.5 cycles
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (<95% dip in Ut)
for 5 sec
3 A/m
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a typical commercial
or hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the product
requires continued operation during power mains
interruptions, it is recommended that the product be
powered from an uninterruptible power supply or a
battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.
15