6
Dentronix Ultrasonic Cleaner
DDUS60R Recessed model
Tank capacity
4.8 gallons (18.3 liters)
Use capacity
4.0 gallons (15.2 liters)
Power
400 watts max
Operating frequency
40 kHz
100-115V Fuse
T8A/250V, 5x20mm
230-240V Fuse
T4A/250V, 5x20mm
Tank dimensions
14.0" x 9.0" x 8.75"
(35.6 x 22.9 x 22.2cm)
Above countertop
17.125" x 12.125" x 2.25"
(43.5 x 30.8 x 5.7cm)
Below countertop
17.125" x 13.0" x 14.0"
(43.5 x 33.0 x 35.6cm)
Electronic enclosure
17.0" x 7.5" x 4.5"
(43.2 x 19.0 x 11.4cm)
Complies with the MDD93/42/EEC
ISO 3864, No. B.3.1
Caution (refer to accompanying documents)
ISO 3864, No. B.3.6
Caution, risk of electric shock
C-UL-US listing is for the following catalog number
items only: DDUS60-115, DDUS60R-115
417-IEC-5016-a
Fuse location
X. Maintenance and Service
A. Cleaning
All components should be cleaned each day by wiping with a
soft cloth moistened with a commercially available mineral
deposit remover. Follow by cleaning with isopropyl alcohol or a
mild disinfectant. The stainless steel surfaces of the enclosures
should be cleaned with a stainless steel cleaner or isopropyl
alcohol when needed. Do not use phenolic based cleaners on
the electronics enclosure (DDUS60R only).
CAUTION:
Do not use petroleum based solvents, iodophors or
phenolic based products. (Iodophors and phenolics can stain
the surface of the unit.) Clean up all liquid spills immediately.
B Inspecting the Power Cord
Periodically check the power cord for damage and see that the
connectors at both ends of the cord are fully seated.
C. Changing the Fuse
1. Turn the main power switch to the “off” position.
2. Remove the power cord from the wall outlet first and then
from the power cord receptacle on the rear of the unit.
!
3. The fuse is held by a fuse drawer which is inserted into the
fuse holder. The fuse holder remains fixed to the
DDUS60/DDUS60R housing. The fuse drawer encircles the
fuse and allows the fuse to be removed from the fuse holder.
These parts are located directly next to the power cord
receptacle. To access the fuse in the 115V unit, insert a small,
flat blade screwdriver in the slot on the fuse drawer and pry
the drawer open. To access the fuse on the 230V unit,
depress the center tab on the fuse drawer and pull the
drawer open.
4. Remove the fuse from the fuse drawer.
5. Examine the fuse. If it is blown, replace with a fuse which
matches the fuse type listed under the unit specifications
section of this manual. If fuse appears to be in good condi-
tion and power is not activated when the main power on/off
switch is turned “on,” return the unit to Dentronix.
6. Slide the fuse back into the holder. To secure the fuse drawer,
simply push the drawer closed.
7. Make sure the main power switch is “off.” Plug the power cord
into the power cord receptacle on the rear of the unit. Then
plug the power cord into the wall receptacle.
D. Service
This unit must be serviced by an authorized Dentronix Service
Center and should always be shipped complete with tank
assembly and electrical control unit.
XI. Limited Warranty
This product has been carefully manufactured and has met strin-
gent quality assurance requirements. If, in normal use, it is found
defective in material or workmanship within a period of one year
from date of purchase, it will be repaired or at Dentronix’s option,
replaced. Tampering with any of the components, misuse, negli-
gence, alteration, water damage, accident, or lack of reasonable
or proper maintenance and care, will void this warranty.
THE FOREGOING IS IN LIEU OF ANY EXPRESSED OR
IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION,
ANY WARRANTIES OR MERCHANTABILITY OR FITNESS
FOR USE.
DENTRONIX ASSUMES NO RESPONSIBILITY FOR ANY
INCONVENIENCE, LOSS, INJURY, OR DIRECT, INDIRECT, OR
CONSEQUENTIAL DAMAGE ARISING FROM THE POSSES-
SION OR USE OF THE PRODUCT.
Any claim for damage or breakage in transit should be made at
once against the carrier. If factory service is required, be certain
to properly pack your equipment and return prepaid and insured
to the factory.
Dentronix is a division of Coltène/Whaledent Inc.