2 Safety instructions
2.5 Instruction of practice personnel
Dentsply Sirona
Operating Instructions DAC Universal
16
6681006 D3708
D3708.201.01.01.02 04.2019
CAUTION
Risk of burns!
The drain hose, filter housing, drain filter and drain tank are very hot
during the reprocessing cycle. Switch the DAC Universal off and let it
cool down (about 15 minutes) before emptying and cleaning the drain
tank or cleaning or replacing the drain filter.
CAUTION
Risk of burns!
Note that the temperature of the drain water is hot (approx. 100 °C).
CAUTION
Risk of burns!
The drain tank may contain hot water. Make sure that you do not tip it
over.
2.5
Instruction of practice personnel
WARNING
The personnel responsible for the operation and maintenance of the
DAC Universal must have received instruction on the safe operation
of the DAC Universal from a service technician trained by the
manufacturer. Following the initial instruction from the service
technician, a party must be named in the practice who is responsible
for supporting and training the personnel charged with operating and
maintaining the DAC Universal.
2.6
Installation requirements
CAUTION
Risk of crushing
The overall height of the open DAC UNIVERSAL is 59 cm.
The recommended minimum height for installing the DAC Universal is
70 cm, in order to prevent crushing and injuries when removing the lid
and hitting any hanging cabinets.
WARNING
Risk of contamination!
Please note that the DAC Universal must have warmed up to room
temperature before being operated. The DAC Universal may NOT be
installed in operating state in rooms with temperatures below 0°C,
otherwise ice may form.
WARNING
Risk of contamination!
After installation, the DAC Universal must be acclimatized before use
so the cleaning result is not affected.