2 Safety instructions
2.6 Installation requirements
6681006 D3708
D3708.201.01.01.02 04.2019
17
WARNING
Risk of contamination!
The DAC Universal my only be operated at ambient temperatures of
+16°C to +30°C.
WARNING
Risk of contamination!
The DAC Universal must not be installed and operated in direct
sunlight as this would influence the cleaning result.
WARNING
Risk of contamination!
The DAC Universal may only be operated up to an altitude of 2,000 m
above sea level.
WARNING
Risk of contamination!
The relative humidity must be below 80%.
WARNING
Risk of crushing!
The DAC Universal must be placed on a stable installation site.
WARNING
Risk of contamination!
We recommend that the compressed air supply that is connected to
the DAC Universal complies with the standard ISO 7494-2.
WARNING
Risk of contamination!
Only ISO 7494-2 dry compressed air may be used.
WARNING
Risk of contamination!
The sterile filter enclosed must be installed properly. Only sterile
filters approved by Dentsply Sirona may be used.
DANGER
Electrical shock!
Only the power plugs supplied or original spare parts may be used.
These meet the following requirements: 125 V/min 15 A or 250 V/min
10 A.