4 Operation
4.2 Reprocessing with the DAC UNIVERSAL
Dentsply Sirona
Operating Instructions DAC Universal
66
6681006 D3708
D3708.201.01.01.02 04.2019
4.2.2
Filling the water tank
NOTE
Remove the water tank from the DAC Universal before filling it up.
This prevents water from spilling into the inside of the DAC Universal
housing.
IMPORTANT
The water tank can be removed at any time and without any problems
as long as no reprocessing cycle is running.
If a cycle is running, an error message is generated when the water
tank is removed.
Removing the water tank / filling the water tank / reinserting the
water tank
ü
The DAC Universal is switched on.
ü
Demineralized / distilled water (water conductivity < 3.0 µS/cm) is
available.
1.
Open the front flap of the DAC Universal.
2.
Remove the water tank by carefully pulling the water tank up and
out of the DAC Universal by the handle.
3.
Place the water tank on a flat surface.
4.
Remove the lid of the water tank.
5.
Fill the demineralized / distilled water (water conductivity < 3.0 µS/
cm) in the water tank (min. 1.0 liters, max. 2 liters)
6.
Put the lid back on the water tank.
7.
Insert the water tank back into the DAC Universal.
IMPORTANT
When inserting make sure that the guide hooks on the top rear side of
the water tank are inserted correctly into the housing.
4.2.3
Removing the drain tank
Frequency: As soon as it is full
WARNING
Risk of contamination!
Do not start a cycle if the DAC Universal is not connected to the drain
tank or a siphon or if the drainage system installation was not
performed correctly.
CAUTION
Risk of burns!
The drain hose, filter housing, drain filter and drain tank are very hot
during the reprocessing cycle. Switch the DAC Universal off and let it
cool down (about 15 minutes) before emptying and cleaning the drain
tank or cleaning or replacing the drain filter.
ü
The DAC Universal is switched off and cooled down.
ü
The drain tank is cooled down.