4 Operation
4.2 Reprocessing with the DAC UNIVERSAL
Dentsply Sirona
Operating Instructions DAC Universal
74
6681006 D3708
D3708.201.01.01.02 04.2019
CAUTION
Risk of heat!
When you take items out of the chamber, remember that the chamber
and these items can be very hot.
Ä
There are instruments on the Green lid; the DAC Universal is still
open and ready for a cycle.
Reprocessing
04:00 PM
1.
Press the "
Start“ function key to start the cycle.
Ä
The Green lid of the DAC Universal closes, the DAC Universal
detects the lid and selects the correct program.
Ä
The DAC Universal automatically runs the cleaning and
disinfection cycle.
Ä
The display shows at any time which phase the DAC Universal
is in, see section "Process symbols [→ 39]“.
04:00 PM
214
134°C
0.5 min
Ä
After the selected cycle has ended, the message appears on
the display see image on left.
Ä
The lid moves a little out of the chamber. This indicates that a
cycle has been completed and the instruments have been
reprocessed.
CAUTION
Risk of burns!
Never hold your head directly over the DAC Universal when it opens,
as steam may emerge from the device.
WARNING
Risk of contamination!
After completion of the cycle, make sure that all instruments are still
properly attached to the adapters. If instruments have fallen into the
chamber during the cycle or if an instrument is no longer properly
attached after the end of a cycle, cleaning and disinfection have
failed. In this case, reprocessing of the instruments must be repeated.
2.
Press the "
Confirm" function key.
Ä
The Green lid will open completely.
CAUTION
Risk of crushing!
Make sure that you do not trap your fingers between the opening lid
and an object, e.g., a wall cupboard which is located above.