Step by Step Instructions
21
5 .3 .4 .2 Flow Handpiece Features
FLOW HEATER (HTR)
Removing the HTR initiates GP push-rod retraction.
CUFF SWITCH
Press the raised ridge on the cuff switch to initiate heating, if
not already at set temperature. Heating takes approximately 90
seconds. The heater will turn off after 10 minutes of inactivity to
conserve battery power.
With HTR at set temperature, press and hold to extrude GP.
With no HTR installed, press to interrupt GP push-rod retrac-
tion. Each additional press alternates the push-rod direction, if
necessary for cleaning.
GP PUSH-ROD DIRECTION INDICATOR
Blue:
GP push-rod moving forward
Flashing blue:
GP push-rod retracting
GP LEVEL INDICATOR
Green:
GP level full
Yellow:
GP level low
Red:
GP level critically low/empty
HEATER STATUS INDICATOR
White light off:
Heater off
Flashing white:
Heating up
White:
Set temperature reached and maintained