Cleaning, Disinfection and Sterilization
26
6 .5
Gutta-Percha Cartridges (GP)
Step
Action
A
Prior to using a GP cartridge on a patient, wipe the cannula with isopropyl
alcohol and allow to dry.
GP cartridges are for single patient use only.
Do not immerse GP cartridges into any liquid.
Do not use GP cartridges after the expiration date.
Dispose of GP cartridges in a sharps receptacle.
GP cartridges cannot be reheated or reused.
Reheating or reusing a GP cartridge can result in the cartridge rupturing,
which will permanently damage the handpiece. Reheating GP cartridges is a
misuse of the
Gutta‑Smart™
and voids the device warranty.
Store GP cartridges at room temperature.