Cleaning, Disinfection and Sterilization
24
6
Cleaning, Disinfection and Sterilization
6 .1
Forward
The components of the
Gutta‑Smart™
have not been disinfected or sterilized
when they are delivered. Please perform the necessary steps before every treat-
ment.
You are responsible for the sterility of accessories; always ensure that only validated
methods of cleaning/disinfection and sterilization are used, that regular mainte-
nance and inspections are performed on the equipment (disinfector, sterilizer) and
that the validated parameters are maintained in every cycle.
Comply with your national guidelines, standards and requirements for cleaning,
disinfection and sterilization. Observe the concentration of the cleaning or disin-
fecting solution stated by the manufacturer.
Failure to adhere to the following instructions or blatant misuse of the device and/
or related components will be considered abuse of the device and will void all
warranty responsibilities.
6 .2
Base Station
Step
Action
A
Clean the exterior of the base station by wiping with a soft cloth moistened
with mild non-chlorinated detergent or disinfecting solution.
Do not immerse the device in any fluid.
6 .3
Handpieces
Step
Action
A
Remove the EHP or TRT from the pack handpiece.
B
Insert the front-end sealing cap firmly into the end of the pack handpiece.
C
Remove the HTR from the flow handpiece.
D
Install the front-end sealing cap firmly onto the end of the flow handpiece.