Dentsply Sirona
Operating Instructions Primescan Connect
2 Safety instructions
2.2 Electrostatic charge
67 97 216 D3801
D3801.201.01.03.02 2022-09
15
2.1.6
Maintenance and repair
As manufacturers of dental instruments and laboratory equipment, we
can assume responsibility for the safety properties of the unit only if the
following points are observed:
● The maintenance and repair of this unit may be performed only by
Dentsply Sirona or by agencies authorized by Dentsply Sirona.
● Components which have failed and influence the safety of the unit
must be replaced with original (OEM) spare parts.
● Only original cables may be used, so that EMC requirements are
met.
Please request a certificate whenever you have such work performed. It
should include:
● The type and scope of work.
● Any changes made in the rated parameters or working range.
● Date, name of company and signature.
2.1.7
Modifications to the product
Modifications to this product which might affect the safety of the system
owner, patients or other persons are prohibited by law!
2.1.8
Accessories
In order to ensure product safety, this device may be operated only with
original Dentsply Sirona accessories or third-party accessories
expressly approved by Dentsply Sirona. In particular, only the power
cable also supplied or the corresponding original spare part may be
used with the unit. The user is responsible for any damage resulting
from the use of non-approved accessories.
2.2
Electrostatic charge
2.2.1
ESD warning labels
ESD warning label
CAUTION
Risk of injury or damage to components from electrostatic discharge
For electrical components labeled with an ESD warning label,
observe the following instructions.
➢ Apply the ESD protective measures.
➢ Do not touch connector pins or sockets without applying ESD
protective measures first.
➢ Do not establish any connections between these connectors
without applying ESD protective measures first.
2.2.2
ESD protective measures
ESD
ESD stands for ElectroStatic Discharge.