6 Maintenance
6.1 Cleaning, disinfection, and sterilization
Dentsply Sirona
Operating Instructions Primescan Connect
48
67 97 216 D3801
D3801.201.01.03.02 2022-09
6.1.2
Non-critical surfaces
NOTE
Never use corrosive cleaning agents, wax or solvents.
CAUTION
Risk of infection
There is a risk of infection if surfaces are not regularly disinfected.
➢ Wipe-disinfect the following non-critical contact surfaces after
each treatment:
- Scanner holder
- Scanner body except sleeve
- Coupling box
- Medical power supply unit
- Scanner cable and all connection cables.
1.
Use a new cotton gauze moistened with one of the cleaning agents
listed in the "Cleaning agents and disinfectants [→ 47]" section to
disinfect the contact surfaces of the product described above.
2.
Dispose of the absorbent gauze.
Please ensure that no colored cloths are used for this as these can
result in discoloration of the surfaces, e.g. when used in combination
with disinfectants.
Protection against medicaments
Due to their high concentrations and the substances they contain, many
medicaments can dissolve, etch, bleach or discolor surfaces.
NOTE
The only way to prevent damage is to wipe off medicaments
immediately with a damp cloth and a cleaning agent!