6 Maintenance
6.1 Cleaning, disinfection, and sterilization
67 97 216 D3801
D3801.201.01.03.02 2022-09
59
6.1.3.8
Hot air sterilization of the window sleeve
NOTE
Hot air sterilization and high-level disinfection must not be combined.
The process for hot air sterilization is as follows:
1.
Carry out a preliminary cleaning process (see "Preliminary cleaning
of the sleeve [→ 52]").
2.
Remove the sleeve from the scanner body (see "Removing the
sleeve [→ 52]") section.
3.
Place the white protective sleeve on the scanner head and place
the scanner body in the scanner holder.
4.
The sleeve can be sterilized using hot air (180°C, 30 min). Please
ask your dealer for deals on hot air sterilizers. Place the sleeves in
the hot air sterilizer and follow the manufacturer's instructions.
5.
Store the sleeve in such a way that it is protected from
contamination until the next use.
6.
Carefully re-attach the sleeve and allow it to lock in place (see
section "Attaching the sleeve [→ 52]"). Hold the scanner in other
places, not just on the sleeve in order to prevent it from falling.