6 Maintenance
6.1 Cleaning, disinfection, and sterilization
Dentsply Sirona
Operating Instructions Primescan Connect
62
67 97 216 D3801
D3801.201.01.03.02 2022-09
12. Slide the sterilized sleeve that has been fitted with the disposable
window onto the scanner body as far as it will go, so that it is
securely positioned on the Primescan scanner. To prevent the
Primescan scanner from falling, hold the Primescan scanner in
other places, not just on the sleeve.
13. Position the Primescan scanner in the holder, so that it can warm
up briefly prior to the intraoral exposure. Make sure that the holder
has been wipe-disinfected.
The following autoclaving processes have been tested and approved for
treatment of the autoclavable sleeves:
Gravity
● 121 °C, 30 min holding time, wrapped
● 132 °C, 15 min holding time, wrapped
● 134 °C, 10 min holding time, wrapped
● 135 °C, 10 min holding time, wrapped
● 134 °C, 3 min holding time, unwrapped
● 135 °C, 3 min holding time, unwrapped
Fractioned pre-vacuum
● 132 °C, 3x fractioned pre-vacuum, 4 min holding time, wrapped
● 134 °C, 3x fractioned pre-vacuum, 3 min holding time, wrapped
● 135 °C, 3x fractioned pre-vacuum, 3 min holding time, wrapped