6 Maintenance
6.1 Cleaning, disinfection, and sterilization
Dentsply Sirona
Operating Instructions Primescan Connect
52
67 97 216 D3801
D3801.201.01.03.02 2022-09
6.1.3.3
Removing the sleeve
If it is necessary to remove the sleeve, proceed as follows:
1.
Grip the sleeve in the marked area.
NOTE
There is a risk of damaging the scanner window or the sleeve window
if the sleeve is not pushed straight toward the front.
➢ Push the sleeve straight toward the front; do not tilt it.
2.
Remove the sleeve from the scanner body.
6.1.3.4
Attaching the sleeve
NOTE
Risk of damage to the windows
There is a risk of damaging the scanner window or the sleeve window
if the sleeve is not pushed in a straight way.
➢ The sleeve must not come into contact with the scanner window.
➢ Push the sleeve straight toward the scanner body; do not tilt it.
➢ Carefully refit the sleeve up to the stop.
6.1.3.5
Preliminary cleaning of the sleeve
Clean the scanner immediately after use as follows:
1.
Carefully wipe the sleeve while it is on the scanner body so that no
dirt whatsoever can remain stuck and can harden on the surface of
the sleeve. For this purpose, use the following:
- a dampened wiping cloth (see "Cleaning agents [→ 47]"),
- an absorbent cotton gauze or a lint-free cloth, dipped in cleaning
agent (see "Cleaning agents [→ 47]").
2.
Wipe down the sleeve afterwards again with the absorbent cotton
gauze dipped in drinking water.
3.
Next dry the sleeve using a lint-free cloth.