background image

Safety Precautions

Safety Precautions (cont.)

• This appliance shall not be used by children. Keep the appliance and its cord out of reach of 

children.

• Please store the ultrasonic cleaner where it is not reachable by children.

• Danger to children! Danger for death through suffocation! Keep the packaging material away 

from children.

• Danger of electrical shock! Never immerse the device or the power cord in water or other liquid. 

• Danger of electrical shock! Never touch the power plug with wet hands, especially when 

inserting or removing the plug.

• Danger of electrical shock! If the unit has fallen into water during operation, do not touch the 

unit.  Remove the power plug from the socket first. 

• Never operate the unit unattended.

• Danger of electrical shock! Do not spray water or liquid over the device.

• Follow the manual to operate the device.

• Do not use components unapproved by the manufacturer.

• When removing the power cord from the socket, grab the power plug not the cord.

• To protect the power cord from damage, do not cause it to get caught by things such as a 

cupboard door or a chair leg; do not drag across a hot surface.

• If there is damage to the power plug, cord, housing, or other parts of the device, do not use  

the device. Contact your dental dealer for a replacement.

• Do not disassemble the device, except by a repair professional.

• If the unit is damaged, non-operational or has fallen into water, contact a qualified service 

provider. 

• Remove the power plug from the socket:

  - if malfunction occurs
  - before cleaning the device
  - if the device is not going to be used for prolonged period
  - after each use (recommended)

• The installation of an earth leakage circuit breaker with a rated tripping current of no more than 

30 mA provides further protection against an electrical shock. The installation should only be 
carried out by a trained electrician. 

Keep away from children!

To prevent life-threatening electrical shock, please observe the following: 

Important Safety Instructions

• Never block the vents on the device. Keep the vents free from lint, hair and other materials. Do 

not place the device on a soft surface where the vents could be blocked. Observe the other 
warnings on the previous page.

•     If the supply cord is damaged, it must be replaced by your dental dealer or its service agent or 

a similarly qualified person in order to avoid a hazard.

Other observations:

• Do not operate the product without filling the tank with water. Running dry will damage the unit.

• Make sure the water level stays below the MAX line after loading the instruments.

 • Do not plug in the power cord before adding water to the tank. Do not fill the tank above the 

Max line with cassettes or trays in place to avoid overspill. 

• Do not use solution containing abrasive substances or strong corrosive chemical solution not 

recommended by the manufacturer or the supplier. Use a recommended cleaning additive such 
as Resurge® Instrument Cleaner.

•  Place the device on a dry and flat surface for operation. 

•  When the device is subjected to severe electromagnetic interference, it may malfunction, stop 

operating or lose control functions. If this happens, unplug the power cord then reinsert it to 
restart the device.

•  Read all the instructions before using the appliance. 

•  Only use attachments recommended or sold by the manufacturer. 

•  Do not use outdoors. 

•  To disconnect, turn the power switch on the back of the unit to the off (0) position, then 

remove plug from outlet. Do not unplug by pulling on cord. To unplug, grasp the plug, not the 
cord. Unplug from outlet when not in use and before servicing or cleaning. 

•  To reduce the risk of electrical shock, do not put the appliance in water or other liquid. Do not 

place or store appliance where it can fall or be pulled into a tub or sink. 

•  All servicing of this product, including transducer replacement, is to be conducted by qualified 

service personnel. 

•  Do not operate any appliance with a damaged cord or plug, or after the appliance 

malfunctions or is dropped or damaged in any manner. 

•  Return appliance to the nearest authorized service facility for examination, repair, or electrical 

or mechanical adjustment. 

To prevent fire hazards, please observe the following:

When using electric appliances, basic precautions should always be followed, 
including the following: 

1

2

Summary of Contents for Resurge

Page 1: ...efficiency not susceptible to interference more durable Multiple circuit protectors When overloaded or improperly used the protectors shut down the power to certain areas to protect the machine Moisture proofed Printed Circuit Board with industrial IC PCBs are placed vertically for better moisture proof performance Better anti interference Three cooling fans Two for cooling of PCB boards one for c...

Page 2: ...e Control panel with capacitive sensing technology durable reliable resistant to water and harsh chemicals Electrical circuit features 1 Three transducers each has an independent PCB control The output power is increased by 40 compared to less expensive control method that uses one PCB to control multiple transducers 2 Multiple circuit protectors a Thermal switch to prevent running dry If temperat...

Page 3: ...lint hair and other materials Do not place the device on a soft surface where the vents could be blocked Observe the other warnings on the previous page If the supply cord is damaged it must be replaced by your dental dealer or its service agent or a similarly qualified person in order to avoid a hazard Other observations Do not operate the product without filling the tank with water Running dry w...

Page 4: ... ready under normal cleaning mode Press Start Stop key twice and the cleaning starts The cleaning stops when the timer counts down to 00 00 If the unit needs to be stopped before the timer runs out touch Start Stop key twice Multistage Thermometer Solid green light indicates that the water temperature is 0 30 of the set temperature Solid orange light indicates that the water temperature is 30 60 o...

Page 5: ...e the gases from the solution to maximize the cavitation energy If you do not degas the cavitation energy can be blocked by the gases making the cavitation energy less effective Adding Instruments a Put items in cassettes or stainless steel basket or trays provided b Lay the cassettes or trays in the metal rack c Lower the rack into the cleaning tank d Confirm that the water level is between MIN a...

Page 6: ...n Guidelines Quick Start cont Operation Guidelines Quick Start cont Notes on Operations This unit is equipped with advanced power off memory capability It memorizes the last timer setting once the cleaning cycle is started a 5 seconds after the timer counts down to 00 00 the last timer setting will reappear Press the Start Stop key twice it will repeat the last cycle with the same timer setting b ...

Page 7: ...r the housing Use a towel to wipe the tank and the housing dry Do not expose the unit under direct sunshine for a long time Keep the unit in a dry cool and ventilated area To clean the outside of the unit use a non abrasive cleaner and soft cloth Do not add cold water when unit is hot 4 5 6 7 Optional Accessories Stainless Steel Rack and Tray 9 10 Double beaker holding set Indirect tank removable ...

Page 8: ...11 12 PAGE INTENTIONALLY BLANK PAGE INTENTIONALLY BLANK ...

Page 9: ...x 6 0 L 1 6 gallon US Min 4 3 L 1 1 gallon US Tank Size 33 1x23 8x10 5 cm 13 0inx9 4inx4 1in Longest Item Fit Inside Tank 37 O cm 14 4in Power Supply 380W AC 100 12ov 60Hz 410W AC 220 240V SOHz 310W AC 100V 50 60Hz Digital Timer Settings Drainage Drainage valve Ultrasonic Frequency 35 000 Hz Tank Material Stainless Steel SUS304 Housing Material ABS Net Weight 9 5 kg 21 1 lb Gross Weight 10 9 kg 24...

Reviews: