7 Replacing components
Dentsply Sirona
7.1 Removing the cover
Service Manual Sinius / Sinius CS / Sinius TS
7.1
Removing the cover
7.1.1
Patient chair
7.1.1.1
Removing the chair covers
Removing the connection box cover
1. Open the connection box cover (A) with a screwdriver.
2. Pull off the connection box cover toward the front.
3. Loosen the 3 screws (C) and remove the cover (B).
Removing the rear covers from the base of the chair
NOTICE
Prying or twisting cover panels causes damage
Do not open the cover by prying it with or twisting the screwdriver! This
can damage the covers.
➢ Push the screwdriver only in the axial direction and raise the cover.
C
A
B
B