63 69 149 D3561
D3561
.
076.01.07
.
02 02.2017
127
Dentsply Sirona
7 Replacing components
Service Manual Sinius / Sinius CS / Sinius TS
7.3 Important information when replacing boards
båÖäáëÜ
7.3
Important information when replacing boards
7.3.1
Checking the software for the entire unit
Before replacing boards
Use the "DUMC" application to perform a data backup before replacing
the following boards.
● NAJm
● HSA
● NWEm
● NHE / NHEm
● HAP2 / HAP1
● NLV
● XLV
Tip: DUMC = (Dental Unit Management Console, see Service support
Following replacement
Always check the global software using the DUMC after replacing boards.
Also update the global software if necessary.
1. Back up all data.
2. Replace the boards.
3. Perform a firmware download.